An ion selective electrode (ISE) turns the activity of one ion in water into a voltage. Three parts do the work: a membrane that responds to the target ion, a reference electrode that holds a steady potential, and a high-impedance voltmeter between them.
What an ion selective electrode measures
The membrane develops a potential that depends on the activity of the target ion on its sample side. Activity is the effective concentration once the other dissolved salts are accounted for, and the two move apart as ionic strength rises.
The reading is reported in mg/L because the calibration standards are labelled in mg/L. That holds while the sample has roughly the same ionic strength as the standards. In the laboratory this is forced by adding a high concentration of an inert electrolyte, an ionic strength adjuster, to every standard and sample, as described in the potentiometry chapter of Harvey’s Analytical Chemistry 2.1. A probe installed in a tank or pipe measures the water as it is, so the standards should resemble that water.
The membrane and the reference
The membrane sets which ion the electrode sees. The same chapter describes the main families:
- Glass membranes, used for pH. A pH electrode is an ion selective electrode for hydrogen ions, and our pH sensor calibration guide covers it separately.
- Solid-state membranes, made from a crystal or pressed salt. A fluoride electrode uses a single crystal of lanthanum fluoride.
- Liquid and polymer membranes, which carry an ionophore, a complexing agent that binds the target ion selectively.
- Gas-sensing electrodes, where a gas-permeable membrane sits in front of an internal electrode. The dissolved CO2 electrode is one of these.
The reference electrode completes the circuit, and its potential has to stay constant whatever the sample contains.
The Nernst slope: 59.16 mV and 29.58 mV per decade
At 25 °C the cell potential follows E = K + (0.05916 / z) × log(a), where z is the charge of the ion and a is its activity. For a monovalent ion such as nitrate, potassium or chloride the ideal slope is 59.16 mV for each tenfold change in activity. For a divalent ion such as calcium it is half of that, 29.58 mV.
The response is logarithmic. Going from 1 to 10 mg/L moves the voltage by as much as going from 100 to 1000 mg/L.
Small voltage errors become percentage errors. Working the equation backwards, an uncertainty of 1 mV corresponds to about 4% of the reading for a monovalent ion and about 8% for a divalent ion. Harvey gives typical accuracies of 1 to 5% for monovalent ions and 2 to 10% for divalent ions.
Working electrodes rarely reach the ideal figure. Our datasheets state 56 ± 3 mV for the nitrite, potassium and chloride electrodes and 26 ± 2 mV at 25 °C for the calcium electrode. The slope is therefore measured during calibration, never assumed.
Selectivity and interfering ions
No membrane responds to one ion only. The selectivity coefficient states how strongly an interfering ion counts compared with the target. At 1.00 the membrane cannot tell them apart. The interferent adds its weighted activity to that of the target inside the logarithm, so the reading comes out high.
On the NO3-100 datasheet, chloride has a selectivity coefficient of 0.0056, and the chloride concentration has to stay below 10 times the nitrate concentration. A conditioning solution can remove an interference in the laboratory. That is not practical for a continuously installed probe, so check the water composition before choosing the electrode. Seawater carries far more chloride than nitrate, which makes the nitrate electrode a freshwater method. Our nitrate monitoring guide covers safe levels and sensor choice in aquaculture.
Each electrode also has a sample pH window, listed in the table below.
Calibration with standards across decades
Calibration fits the two unknowns in the equation, the constant K and the slope. That needs at least two standards, and the useful arrangement is:
- Choose standards a decade apart that bracket the expected reading, for example 10 and 100 mg/L for a sample near 40 mg/L.
- Rinse between standards and let the reading settle before accepting each point.
- Keep standards and sample at the same temperature.
Temperature
The 59.16 figure is for 25 °C. The slope is proportional to absolute temperature, and the constant K shifts with temperature too, so a calibration made in a warm room does not transfer exactly to cold water. Temperature compensation corrects the slope term.
The limits at low concentration
Two things set the floor of an ISE measurement. Near the bottom of its range the response flattens, because the weighted contribution of interfering ions becomes comparable to the target. And an accuracy stated as a percentage of full scale is a fixed band in mg/L, which is a large share of a small reading. The NO3-100 states ±5% of a 1000 mg/L full scale and recommends use at 30 mg/L and above. The reading is nitrate ion, so divide by 4.43 to express it as nitrate-nitrogen.
Lab electrode or digital probe
A laboratory ISE is a passive element. Its output impedance is tens of megohms (20 to 50 MΩ on our datasheets), so it needs a dedicated meter and a short shielded cable, and the calibration lives in the meter.
A digital probe carries the amplifier, the temperature sensor and a microprocessor in the body. It stores its own calibration and sends a finished concentration over RS485, so the cable can be long and the receiving end can be any logger, PLC or microcontroller that speaks Modbus RTU. Our guide to sensor protocols explains the bus.
Agrinovo ion-selective probes
Each probe below is a digital ISE with RS485 Modbus RTU output, 0.01 resolution, a built-in NTC10K thermistor for automatic or manual temperature compensation, and a 12 to 24 V DC supply. Accuracy is ±5% of full scale on all of them.
| Probe | Ion | Range | Sample pH | Interfering ions listed |
|---|---|---|---|---|
| NO3-100 | Nitrate, NO₃⁻ | 0 to 1000 mg/L (customizable), recommended 30 mg/L and above | Cl⁻, ClO₄⁻, I⁻, CN⁻, BF₄⁻, NO₂⁻ | |
| NH4-100 | Ammonium, stated on the datasheet as ammonium nitrogen (NH₄⁺-N) | 0.2 to 1000 ppm | ||
| NO2-100 | Nitrite, NO₂⁻ | 0.2 to 1000 ppm (customizable up to 2200 ppm) | 2 to 12, keep the sample between 4.5 and 8 | Acetate, fluoride, chloride, nitrate, sulfate |
| K-100 | Potassium, K⁺ | 0.7 to 1000 ppm (customizable up to 98,000 ppm) | 2 to 12 | Cs⁺, NH₄⁺ |
| CA-100 | Calcium, Ca²⁺ | 0.2 to 1000 ppm (customizable up to 40,000 ppm) | 3 to 10 | Pb²⁺, Hg²⁺, Cu²⁺, Ni²⁺ |
| CL-100 | Chloride, Cl⁻ | 1 to 1000 ppm (customizable up to 35,500 ppm) | 2 to 12 | S²⁻, I⁻, CN⁻, Br⁻ |
| F-100 | Fluoride, F⁻ | 0 to 1000 mg/L (customizable) | ||
| CU-100 | Copper, Cu²⁺ | 0.064 to 1000 ppm (customizable up to 6400 ppm) | 2 to 12 | Hg²⁺, Ag⁺, Fe²⁺, Br⁻, Cl⁻ |
| PB-100 | Lead, Pb²⁺ | 0.2 to 1000 ppm (customizable up to 21,000 ppm) | 0 to 12 | Ag⁺, Hg²⁺, Cu²⁺, high levels of Cd²⁺, Fe²⁺ |
| AG-100 | Silver, Ag⁺ | 0.1 to 1000 ppm (customizable up to 100,000 ppm) | 2 to 8 | Hg²⁺ |
For choosing between the ammonium, nitrite and nitrate probes in fish culture, see ammonia, nitrite and nitrate sensors for aquaculture.
Reading and calibrating over RS485
- Measurement. Ion concentration, the raw electrode signal in mV and temperature are read in one request, as integers (function 0x04) or 32-bit floats (function 0x03).
- Valence. Register 0x0023 is set to 1 for a monovalent ion or 2 for a divalent ion before calibration, to match the charge of the ion.
- Calibration. The electrode needs two or more standards before it reports a concentration and stores up to five points. The standards are 0.1, 1, 10, 100 and 1000 ppm, each with its own code written to register 0x0043 while the electrode sits in that solution. Reading the same register returns the status: 0 for success, 1 while calibrating, 2 to 4 for a failed point. The NO3-100 datasheet recommends three points at 10, 100 and 1000 ppm.
- Electrode efficiency. Each successful calibration recalculates electrode efficiency, readable from register 0x001C, with an accepted range of 70.0 to 130.0%. A point that falls outside the limits returns status 4. Log the value at each calibration: a membrane that has aged, dried out or fouled gives fewer millivolts per decade.
The full register maps and example frames are on the developer pages for the NO3-100, NH4-100 and AG-100. For quotes and purchase orders through a university purchasing office, see purchasing for research.