NH4-100 with Arduino, ESP32, Raspberry Pi and Python
How to wire the NH4-100 Ammonium Ion-Selective Electrode to your own microcontroller or computer and read it over RS485 Modbus RTU. The tables are the ones in the datasheet, and both code examples are run against the datasheet example frame before this page is published.
Connecting the NH4-100
The probe is an RS485 device, so it needs a transceiver between it and your board. On a computer or a Raspberry Pi that is a USB to RS485 adapter. On a microcontroller it is a MAX485-type module on a hardware serial port. Power the probe from a supply within the voltage range in the table and join its ground to the ground of the transceiver.
| Wire Color | Function | Description |
|---|---|---|
| Red | V+ | DC Power Supply (12-24V DC) |
| Black | GND | DC Power Ground |
| Yellow | RS485-A | Data T/R+ |
| White | RS485-B | Data T/R− |
Python example (Raspberry Pi, Linux, macOS, Windows)
The script needs only pyserial. It builds the Modbus request itself, checks the CRC of the reply and prints one reading every five seconds. On Windows pass the COM port as the argument.
#!/usr/bin/env python3
"""Read the Agrinovo NH4-100 over RS485 Modbus RTU.
Works on a Raspberry Pi or any computer with a USB to RS485 adapter.
Install the serial library once: pip install pyserial
Run: python3 nh4_100.py /dev/ttyUSB0
"""
import struct
import sys
import time
import serial
PORT = sys.argv[1] if len(sys.argv) > 1 else "/dev/ttyUSB0"
ADDRESS = 0x01 # the probe's Modbus address
def crc16(frame: bytes) -> bytes:
crc = 0xFFFF
for byte in frame:
crc ^= byte
for _ in range(8):
crc = (crc >> 1) ^ 0xA001 if crc & 1 else crc >> 1
return struct.pack("<H", crc)
def read_registers(port, address, function, start, count):
request = struct.pack(">BBHH", address, function, start, count)
port.reset_input_buffer()
port.write(request + crc16(request))
reply = port.read(5 + 2 * count)
if len(reply) >= 5 and reply[1] == function | 0x80:
raise IOError(f"probe returned Modbus error {reply[2]}")
if len(reply) != 5 + 2 * count or crc16(reply[:-2]) != reply[-2:]:
raise IOError("no valid reply from the probe")
return reply[3:-2]
def read_probe(port):
data = read_registers(port, ADDRESS, 0x03, 0x0000, 10)
return {
"ammonium_mg_l": struct.unpack(">f", data[2:4] + data[0:2])[0],
"electrode_mv": struct.unpack(">f", data[6:8] + data[4:6])[0],
"temperature_c": struct.unpack(">f", data[18:20] + data[16:18])[0],
}
if __name__ == "__main__":
with serial.Serial(PORT, 9600, bytesize=8, parity="N", stopbits=1, timeout=1) as port:
while True:
print(read_probe(port))
time.sleep(5) Arduino example (ESP32, Mega and other boards with a second serial port)
The sketch uses no library. On an ESP32 it uses Serial2 on the pins set at the top; on AVR boards it uses Serial1. DE and RE of the transceiver go to one output pin.
// Agrinovo NH4-100 over RS485 Modbus RTU.
// For a board with a spare hardware serial port (ESP32, Mega, Leonardo, Due)
// and a MAX485-type transceiver. Power the probe from its own supply, as the
// wiring table says, and join the grounds.
const uint8_t ADDRESS = 0x01; // the probe's Modbus address
const int DE_PIN = 4; // transceiver DE and RE tied together
const int RX_PIN = 16; // ESP32 only
const int TX_PIN = 17; // ESP32 only
#if defined(ESP32)
#define RS485 Serial2
#else
#define RS485 Serial1
#endif
uint16_t crc16(const uint8_t *frame, size_t length) {
uint16_t crc = 0xFFFF;
for (size_t i = 0; i < length; i++) {
crc ^= frame[i];
for (int bit = 0; bit < 8; bit++) {
crc = (crc & 1) ? (crc >> 1) ^ 0xA001 : crc >> 1;
}
}
return crc;
}
// Builds a float from its four bytes, most significant first.
float float32(uint8_t b3, uint8_t b2, uint8_t b1, uint8_t b0) {
uint32_t bits = ((uint32_t)b3 << 24) | ((uint32_t)b2 << 16) | ((uint32_t)b1 << 8) | b0;
float value;
memcpy(&value, &bits, sizeof value);
return value;
}
// Reads `count` registers into `data`. Returns false when the probe does not
// answer or the checksum is wrong.
bool readRegisters(uint8_t function, uint16_t start, uint16_t count, uint8_t *data) {
uint8_t request[8] = {ADDRESS, function, (uint8_t)(start >> 8), (uint8_t)start,
(uint8_t)(count >> 8), (uint8_t)count, 0, 0};
uint16_t crc = crc16(request, 6);
request[6] = crc & 0xFF;
request[7] = crc >> 8;
while (RS485.available()) RS485.read();
digitalWrite(DE_PIN, HIGH);
RS485.write(request, 8);
RS485.flush();
digitalWrite(DE_PIN, LOW);
uint8_t reply[5 + 2 * 16];
size_t expected = 5 + 2 * count;
if (RS485.readBytes(reply, expected) != expected) return false;
uint16_t received = reply[expected - 2] | ((uint16_t)reply[expected - 1] << 8);
if (reply[1] != function || crc16(reply, expected - 2) != received) return false;
memcpy(data, reply + 3, 2 * count);
return true;
}
void setup() {
Serial.begin(115200);
pinMode(DE_PIN, OUTPUT);
digitalWrite(DE_PIN, LOW);
#if defined(ESP32)
RS485.begin(9600, SERIAL_8N1, RX_PIN, TX_PIN);
#else
RS485.begin(9600);
#endif
RS485.setTimeout(1000);
}
void loop() {
uint8_t data[2 * 10];
if (!readRegisters(0x03, 0x0000, 10, data)) {
Serial.println("no reply");
delay(5000);
return;
}
Serial.print("ammonium_mg_l=");
Serial.print(float32(data[2], data[3], data[0], data[1]), 2);
Serial.print(" electrode_mv=");
Serial.print(float32(data[6], data[7], data[4], data[5]), 1);
Serial.print(" temperature_c=");
Serial.print(float32(data[18], data[19], data[16], data[17]), 1);
Serial.println();
delay(5000);
} What the examples read
Both examples send one request, function 0x03, 10 registers from 0x0000, and decode these values:
| Value | Unit | Register | Encoding |
|---|---|---|---|
ammonium_mg_l | mg/L | 0x0000 | 32-bit float in two registers, low word first |
electrode_mv | mV | 0x0002 | 32-bit float in two registers, low word first |
temperature_c | °C | 0x0008 | 32-bit float in two registers, low word first |
Other Modbus tools
The same request works from any Modbus RTU master: pymodbus, minimalmodbus, Node-RED, LabVIEW, a PLC or a third-party data logger with a Modbus port. Use the serial settings, function code and register addresses above.
Logging without writing code
The Omni Genesis controller reads the NH4-100 and sends the readings to the Agrinovo cloud over the cellular network, where they can be exported or read through the cloud API.
Datasheet reference
Specifications
| Parameter | Specification |
|---|---|
| Ion Range | 0.2 to 1000 ppm |
| Ion Resolution | 0.01 ppm |
| Ion Accuracy | ±5% F.S. |
| Temperature Range | −10.0 to 110.0°C |
| Temperature Resolution | 0.1°C |
| Temperature Accuracy | ±0.5°C |
| Temperature Sensor | Thermistor NTC10K |
| Temperature Compensation | Automatic / Manual |
Electrical Characteristics
| Parameter | Specification |
|---|---|
| Supply Voltage | 12-24 VDC |
| Power Consumption | ≤0.5W |
| Isolation | 2500Vrms |
| Housing Material | PPS |
| Mounting Thread | 3/4 inch front and rear threads |
Communication Settings
| Parameter | Value |
|---|---|
| Protocol | Modbus-RTU |
| Baud Rate | 9600 bps (configurable: 1200/2400/4800/9600/19200) |
| Data Bits | 8 |
| Parity | None |
| Stop Bits | 1 |
| Default Address | 0x01 (range 1 to 247) |
Notes:
- All registers are dual-byte (16-bit). Transmitted high byte first, low byte second.
- Negative integers use two’s complement (−1 = 0xFFFF, −2 = 0xFFFE).
- CRC16: Standard Modbus, high byte first, low byte second. CRC bypass code: 0x2A2A.
- Command frame length is fixed at 8 bytes. Frames shorter than 8 bytes after a 0.1s gap are ignored.
Register Map
The sensor has three register groups: Measurement (real-time data), Parameter (configuration), and Information (device status/identity).
Measurement Registers - Integer Mode (Function 0x04)
| Address | Description | Range | Scaling | R/W |
|---|---|---|---|---|
| 0x0000 | Ion Concentration | 0 to 20000 | See decimal register | R |
| 0x0001 | Ion Decimal + Unit | - | See unit table | R |
| 0x0002 | Electrode Signal | −7500 to 7500 | ÷10 → mV | R |
| 0x0003 | Signal Decimal + Unit | - | See unit table | R |
| 0x0008 | Temperature | −100 to 1100 | ÷10 → °C | R |
| 0x0009 | Temp Decimal + Unit | - | See unit table | R |
Over-range indicators: 0x7FFF = above upper limit, 0x8000 = below lower limit.
Measurement Registers - Float Mode (Function 0x03)
Each value occupies 2 registers (float32):
| Address | Description | Range | Unit | R/W |
|---|---|---|---|---|
| 0x0000 to 0x0001 | Ion Concentration | 0.000 to 20000 | ppm | R |
| 0x0002 to 0x0003 | Electrode Signal | −750.0 to 750.0 | mV | R |
| 0x0008 to 0x0009 | Temperature | −10.0 to 110.0 | °C | R |
Parameter Registers (Function 0x03 read / 0x06 write)
| Address | Description | Range | Default | R/W |
|---|---|---|---|---|
| 0x0019 | Calibration Status | BIT1 - BIT5 (see calibration) | - | R/D |
| 0x001C | Electrode Efficiency | 70.0 to 130.0% (int: 700 to 1300) | 100.0% | R/D |
| 0x001D | Calibrated Point Count | 0 to 5 | 0 | R/D |
| 0x001E | Slave Address | 1 to 247 (0xFF = broadcast) | 1 | R/W |
| 0x001F | Baud Rate | 0:1200 · 1:2400 · 2:4800 · 3:9600 · 4:19200 | 3 (9600) | R/W |
| 0x0020 | Temp Compensation Type | 0: Manual · 1: Automatic | 1 (Auto) | R/W/D |
| 0x0021 | Temp Setpoint / Offset | Manual: −10.0 to 110.0°C · Auto: −10.0 to 10.0°C (×10) | 25.0 / 0.0°C | R/W/D |
| 0x0023 | Ion Valence | 0: Not set · 1: Monovalent · 2: Divalent | - | R/W |
R/W/D: D = reset to default by factory restore. Registers without D are unaffected by factory restore.
Information Registers (Function 0x03 read / 0x06 write)
| Address | Description | Values | R/W |
|---|---|---|---|
| 0x0040 | Working Mode | 0x0010: Measurement · 0x0050: Settings · 0x0060: Calibration | R/W |
| 0x0041 | Mode Parameter | See settings/calibration sections | R/W |
| 0x0042 | Work Event | - | R |
| 0x0043 | Calibration Status & Operation | See calibration section | R/W |
| 0x0044 | Device Type | 0x0010 (ION) | R |
| 0x0045 | Device Model | 0x1210 (ION1210) | R |
| 0x0046 | Software Version | BCD coded | R |
| 0x0047 | Hardware Version | BCD coded | R |
| 0x0048 | Serial Number High | BCD coded | R |
| 0x0049 | Serial Number Low | BCD coded | R |
Reading Data
Integer Mode - Read ion concentration, electrode signal, and temperature (Function 0x04)
Read 10 registers from 0x0000:
Request: 01 04 00 00 00 0A 70 0D
Response: 01 04 14 03 EB 02 11 03 EC 01 00 00 00 00 00 00 00 00 00 00 FA 01 0B F1 B7
Decoding:
| Register | Hex | Decimal | Scaling | Result |
|---|---|---|---|---|
| 0x0000 Ion Value | 0x03EB | 1000 | ÷100 (from 0x0001) | 10.00 ppm |
| 0x0002 Electrode Signal | 0x03EC | 1001 | ÷10 (from 0x0003) | 100.1 mV |
| 0x0008 Temperature | 0x00FA | 250 | ÷10 (from 0x0009) | 25.0°C |
Float Mode - Read ion concentration, electrode signal, and temperature (Function 0x03)
Read 10 registers from 0x0000:
Request: 01 03 00 00 00 0A C5 CD
Response: 01 03 14 00 00 41 20 33 33 42 C8 00 00 00 00 00 00 00 00 E3 E8 41 C7 43 0C
Decoding:
| Register | Float Value | Result |
|---|---|---|
| 0x0000 to 0x0001 Ion | 10.00 | 10.00 ppm |
| 0x0002 to 0x0003 Electrode | 100.1 | 100.1 mV |
| 0x0008 to 0x0009 Temperature | 24.986282 | 25.0°C |
Address Configuration
Change Address (0x01 → 0x02)
Request: 01 06 00 1E 00 02 68 0D
Response: 01 06 00 1E 00 02 68 0D
Change Baud Rate (9600 → 2400)
Request: 01 06 00 1F 00 01 79 CC
Response: 01 06 00 1F 00 01 79 CC
Change Ion Valence (→ Divalent)
Request: 01 06 00 23 00 02 F9 C1
Response: 01 06 00 23 00 02 F9 C1
Broadcast Discovery
Use address 0xFF with only one device connected.
Calibration
The ion electrode requires calibration with 2 or more standard solutions before it can calculate ion concentration. Up to 5 calibration points are supported.
Standard Solution Codes
| Concentration | Code | Bit Position |
|---|---|---|
| 0.1 ppm | 0x0002 | BIT1 |
| 1 ppm | 0x0004 | BIT2 |
| 10 ppm | 0x0008 | BIT3 |
| 100 ppm | 0x0010 | BIT4 |
| 1000 ppm | 0x0020 | BIT5 |
Calibration Flow
- Place the electrode in standard solution.
- Write the standard solution code to calibration register (0x0043).
- Wait for completion. Read 0x0043 status:
- 0: Calibration successful (returned to measurement mode)
- 1: Calibrating - read again shortly
- 2: No valid standard value received within 180s (returned to measurement)
- 3: Signal unstable or out of range within 180s (returned to measurement)
- 4: Sensor performance (efficiency/offset) out of limits (returned to measurement)
Calibration Notes
- Each successful calibration calculates electrode efficiency. If out of range, error 4 is returned.
- Writing 0x7FFF to 0x0043 clears all calibration data.
- Calibration status can be read from register 0x0019 and electrode efficiency from 0x001C.
- Ion valence must be configured (0x0023) before calibration if applicable.
Calibration Commands
| Command | Addr | Func | Register | Value | CRC |
|---|---|---|---|---|---|
| Clear all calibration | 01 | 06 | 00 43 | 7F FF | 18 6E |
| Calibrate 0.1 ppm | 01 | 06 | 00 43 | 00 02 | F9 DF |
| Calibrate 1 ppm | 01 | 06 | 00 43 | 00 04 | 79 DD |
| Calibrate 10 ppm | 01 | 06 | 00 43 | 00 08 | 79 D8 |
| Calibrate 100 ppm | 01 | 06 | 00 43 | 00 10 | 79 D2 |
| Calibrate 1000 ppm | 01 | 06 | 00 43 | 00 20 | 79 C6 |
| Query calibration status | 01 | 03 | 00 43 | 00 01 | 75 DE |
Temperature Compensation
Set to Manual Compensation
Request: 01 06 00 20 00 00 88 00
Response: 01 06 00 20 00 00 88 00
When manual: register 0x0021 is the manual temperature setpoint (×10).
Set to Automatic Compensation
Request: 01 06 00 20 00 01 49 C0
Response: 01 06 00 20 00 01 49 C0
When automatic: register 0x0021 is the temperature offset (×10). Default 0.0°C.
Set Temperature Offset to −5.0°C
Request: 01 06 00 21 FF CE 19 A4
Response: 01 06 00 21 FF CE 19 A4
Factory Reset
Factory reset requires two sequential commands:
Step 1 - Enter Settings Mode (write 0x0050 to 0x0040):
Request: 01 06 00 40 00 50 88 22
Response: 01 06 00 40 00 50 88 22
Step 2 - Execute Factory Reset (write 0x7FFF to 0x0041):
Request: 01 06 00 41 7F FF B9 AE
Response: 01 06 00 41 7F FF B9 AE
This clears all calibration data, resets temperature mode to automatic, offset to 0.0°C, and restores default parameter values. The device will restart.
Error Handling
When an invalid command is received, the device responds with the function code + 0x80:
| Error Code | Description |
|---|---|
| 01 | Invalid function code (only 0x03, 0x04, 0x06 supported) |
| 02 | Invalid register address |
| 03 | Invalid register count (exceeds register group boundary) |
| 04 | Invalid modification value (out of range) |
| 05 | CRC error |
| 06 | Write to read-only register |
Common Commands Reference
Read Registers
| Command | Addr | Func | Start | Count | CRC |
|---|---|---|---|---|---|
| Read measurement (integer) | 01 | 04 | 00 00 | 00 0A | 70 0D |
| Read measurement (float) | 01 | 03 | 00 00 | 00 0A | C5 CD |
| Read all parameters | 01 | 03 | 00 19 | 00 0C | 94 08 |
| Read all info registers | 01 | 03 | 00 40 | 00 0A | C4 19 |
Write Registers
| Command | Addr | Func | Register | Value | CRC |
|---|---|---|---|---|---|
| Set address to 0x02 | 01 | 06 | 00 1E | 00 02 | 68 0D |
| Set baud to 2400 | 01 | 06 | 00 1F | 00 01 | 79 CC |
| Temp comp → manual | 01 | 06 | 00 20 | 00 00 | 88 00 |
| Temp comp → auto | 01 | 06 | 00 20 | 00 01 | 49 C0 |
| Temp offset −5.0°C | 01 | 06 | 00 21 | FF CE | 19 A4 |
| Ion valence → monovalent | 01 | 06 | 00 23 | 00 01 | B9 C0 |
| Ion valence → divalent | 01 | 06 | 00 23 | 00 02 | F9 C1 |
Unit Lookup Table
The low byte of “Decimal + Unit” registers (0x0001, 0x0003, 0x0009) contains the unit code:
| Code | Unit | Code | Unit | Code | Unit |
|---|---|---|---|---|---|
| 0x00 | mV | 0x01 | nA | 0x02 | uA |
| 0x03 | mA | 0x04 | Ω | 0x05 | KΩ |
| 0x06 | MΩ | 0x07 | uS | 0x08 | mS |
| 0x09 | S | 0x0A | pH | 0x0B | °C |
| 0x0C | °F | 0x0D | ug/L | 0x0E | mg/L |
| 0x0F | g/L | 0x10 | ppb | 0x11 | ppm |
| 0x12 | ppt | 0x13 | % | 0x14 | mbar |
| 0x15 | bar | 0x16 | mmHg |
The high byte contains the number of decimal places for the value register.
Revision History
| Rev | Date | Changes |
|---|---|---|
| 1.0 | June 2026 | Initial release |
| 1.1 | August 2026 | Layout revision |
| 1.2 | October 2026 | New layout; corrected Modbus example CRC values |