TUR-100 with Arduino, ESP32, Raspberry Pi and Python
How to wire the TUR-100 Stainless Steel Turbidity Sensor to your own microcontroller or computer and read it over RS485 Modbus RTU. The tables are the ones in the datasheet, and both code examples are run against the datasheet example frame before this page is published.
Product page Datasheet (PDF) Datasheet (Markdown) Arduino sketch (.ino) Python script (.py)
Connecting the TUR-100
The probe is an RS485 device, so it needs a transceiver between it and your board. On a computer or a Raspberry Pi that is a USB to RS485 adapter. On a microcontroller it is a MAX485-type module on a hardware serial port. Power the probe from a supply within the voltage range in the table and join its ground to the ground of the transceiver.
| Wire Color | Function | Description |
|---|---|---|
| Red | V+ | Power Supply (12 to 24 V DC) |
| Black | GND | Power Ground |
| Yellow | RS485-A | Data+ |
| Green | RS485-B | Data- |
The wire colours are confirmed by the label on each cable. Wire the sensor with the power off and check every connection before powering up.
Python example (Raspberry Pi, Linux, macOS, Windows)
The script needs only pyserial. It builds the Modbus request itself, checks the CRC of the reply and prints one reading every five seconds. On Windows pass the COM port as the argument.
#!/usr/bin/env python3
"""Read the Agrinovo TUR-100 over RS485 Modbus RTU.
Works on a Raspberry Pi or any computer with a USB to RS485 adapter.
Install the serial library once: pip install pyserial
Run: python3 tur_100.py /dev/ttyUSB0
"""
import struct
import sys
import time
import serial
PORT = sys.argv[1] if len(sys.argv) > 1 else "/dev/ttyUSB0"
ADDRESS = 0x01 # the sensor's Modbus address
def crc16(frame: bytes) -> bytes:
crc = 0xFFFF
for byte in frame:
crc ^= byte
for _ in range(8):
crc = (crc >> 1) ^ 0xA001 if crc & 1 else crc >> 1
return struct.pack("<H", crc)
def read_registers(port, address, function, start, count):
request = struct.pack(">BBHH", address, function, start, count)
port.reset_input_buffer()
port.write(request + crc16(request))
reply = port.read(5 + 2 * count)
if len(reply) >= 5 and reply[1] == function | 0x80:
raise IOError(f"sensor returned Modbus error {reply[2]}")
if len(reply) != 5 + 2 * count or crc16(reply[:-2]) != reply[-2:]:
raise IOError("no valid reply from the sensor")
return reply[3:-2]
def read_probe(port):
data = read_registers(port, ADDRESS, 0x03, 0x0000, 1)
return {
"turbidity": struct.unpack(">H", data[0:2])[0] * 0.1, # NTU
}
if __name__ == "__main__":
with serial.Serial(PORT, 9600, bytesize=8, parity="N", stopbits=1, timeout=1) as port:
while True:
print(read_probe(port))
time.sleep(5) Arduino example (Uno, Mega, ESP32, ESP8266, MKR, Pico)
The sketch uses the Agrinovo library: in the Arduino IDE open Sketch > Include Library > Manage Libraries and install "Agrinovo". The library picks the serial port for your board, and the pins are listed at the top of AgrinovoDefaultPort.h. DE and RE of the transceiver go to one output pin.
// Agrinovo TUR-100 turbidity sensor
// Prints turbidity every 5 seconds.
//
// Sensor cable
// Red supply +, 12-24 V DC
// Black supply ground, joined to the board's ground
// Yellow RS485 A
// Green RS485 B
//
// The board talks to the sensor through an RS485 transceiver (MAX485 or
// similar). AgrinovoDefaultPort.h lists the serial pins used on each board.
//
// https://agrinovo.io/products/sensors/aqua/turbidity/tur-100-turbidity-sensor/
#include <Agrinovo.h>
#include <AgrinovoDefaultPort.h>
const int DE_PIN = 4; // transceiver DE and RE, tied together
AgrinovoBus bus(AGRINOVO_PORT, DE_PIN);
AgrinovoTUR100 sensor(bus); // Modbus address 1
void setup() {
Serial.begin(115200);
agrinovoPortBegin(9600);
bus.begin();
}
void loop() {
if (sensor.read()) {
Serial.print("Turbidity: ");
Serial.print(sensor.turbidity(), 1);
Serial.println(" NTU");
} else {
Serial.println(sensor.errorText());
}
delay(5000);
} What the examples read
Both examples send one request, function 0x03, 1 registers from 0x0000, and decode these values:
| Value | Unit | Register | Encoding |
|---|---|---|---|
| Turbidity | NTU | 0x0000 | Unsigned 16-bit integer, scaled |
Other Modbus tools
The same request works from any Modbus RTU master: pymodbus, minimalmodbus, Node-RED, LabVIEW, a PLC or a third-party data logger with a Modbus port. Use the serial settings, function code and register addresses above.
Logging without writing code
The Omni Genesis controller reads the TUR-100 and sends the readings to the Agrinovo cloud over the cellular network, where they can be exported or read through the cloud API.
Datasheet reference
Specifications
| Parameter | Specification |
|---|---|
| Measurement Principle | 90° infrared scattered light |
| Measurement Ranges | 0 to 10, 0 to 20, 0 to 100, 0 to 400, 0 to 1000 or 0 to 4000 NTU (custom ranges available) |
| Accuracy | ±2% |
| Resolution | 0.1 NTU (register value ÷10) |
| Stability | ≤1 NTU per 24 hours |
| Output | RS485, Modbus-RTU (4 to 20 mA optional) |
| Supply Voltage | 12 to 24 V DC |
| Power Consumption | ≤0.5 W |
| Operating Environment | 0 to 60°C, ≤85% RH |
| Protection | IP68 |
| Housing | Stainless steel |
| Cable | Low-noise cable, runs of 20 m and more |
Communication Settings
| Parameter | Value |
|---|---|
| Protocol | Modbus-RTU |
| Baud Rate | 9600 bps |
| Data Bits | 8 |
| Parity | None |
| Stop Bits | 1 |
| Default Address | 0x01 |
| Broadcast Address | 0xFE (one sensor on the bus only) |
Register Map
| Address | Description | Type | R/W | Notes |
|---|---|---|---|---|
| 0x0000 | Turbidity | Int | R | NTU = value ÷10 |
| 0x0050 | Slave Address | Int | W | New address |
| 0x0053 | Slope | Int | W | Value ÷1000, default 1000 (1.000) |
| 0x0054 | Intercept | Int | W | Default 0 |
Reading Data
Read turbidity (1 register from 0x0000):
Request: 01 03 00 00 00 01 84 0A
Response: 01 03 02 1A 0A 33 23
Decoding:
| Register | Raw | Decimal | Result |
|---|---|---|---|
| Turbidity | 0x1A0A | 6666 | 666.6 NTU |
Address Configuration
Change Address (0x01 to 0x02)
01 06 00 50 00 02 08 1A
If the current address is unknown, send the read or write to the broadcast address 0xFE with only one sensor connected:
FE 03 00 00 00 01 90 05
Calibration
The sensor ships calibrated. A linear correction is applied in the field by writing a slope and an intercept.
Set Slope to 1.200
01 06 00 53 04 B0 7A AF
Set Intercept to 10
01 06 00 54 00 0A 48 1D
Revision History
| Rev | Date | Changes |
|---|---|---|
| 1.0 | October 2026 | Initial release |
Contact
Agrinovo, https://agrinovo.io/products, [email protected], +972 54 688 8148 (direct / WhatsApp). Specifications subject to change without notice.
Questions about wiring or the protocol?
Tell us which board or logger you are using and what you need to measure.