EC-110 with Arduino, ESP32, Raspberry Pi and Python

How to wire the EC-110 Digital Conductivity & Salinity Sensor - 0 to 200 mS/cm to your own microcontroller or computer and read it over RS485 Modbus RTU. The tables are the ones in the datasheet, and both code examples are run against the datasheet example frame before this page is published.

Product page Datasheet (PDF) Datasheet (Markdown) Arduino sketch (.ino) Python script (.py)

EC-110 Digital Conductivity & Salinity Sensor - 0 to 200 mS/cm

Connecting the EC-110

The probe is an RS485 device, so it needs a transceiver between it and your board. On a computer or a Raspberry Pi that is a USB to RS485 adapter. On a microcontroller it is a MAX485-type module on a hardware serial port. Power the probe from a supply within the voltage range in the table and join its ground to the ground of the transceiver.

Wire Color Function Description
Red V+ Power Supply
Black GND Power Ground
Yellow RS485-A Data +
Green RS485-B Data -

Python example (Raspberry Pi, Linux, macOS, Windows)

The script needs only pyserial. It builds the Modbus request itself, checks the CRC of the reply and prints one reading every five seconds. On Windows pass the COM port as the argument.

#!/usr/bin/env python3
"""Read the Agrinovo EC-110 over RS485 Modbus RTU.

Works on a Raspberry Pi or any computer with a USB to RS485 adapter.
Install the serial library once:  pip install pyserial
Run:  python3 ec_110.py /dev/ttyUSB0
"""
import struct
import sys
import time

import serial

PORT = sys.argv[1] if len(sys.argv) > 1 else "/dev/ttyUSB0"
ADDRESS = 0x01  # the sensor's Modbus address


def crc16(frame: bytes) -> bytes:
    crc = 0xFFFF
    for byte in frame:
        crc ^= byte
        for _ in range(8):
            crc = (crc >> 1) ^ 0xA001 if crc & 1 else crc >> 1
    return struct.pack("<H", crc)


def read_registers(port, address, function, start, count):
    request = struct.pack(">BBHH", address, function, start, count)
    port.reset_input_buffer()
    port.write(request + crc16(request))
    reply = port.read(5 + 2 * count)
    if len(reply) >= 5 and reply[1] == function | 0x80:
        raise IOError(f"sensor returned Modbus error {reply[2]}")
    if len(reply) != 5 + 2 * count or crc16(reply[:-2]) != reply[-2:]:
        raise IOError("no valid reply from the sensor")
    return reply[3:-2]


def read_probe(port):
    data = read_registers(port, ADDRESS, 0x03, 0x0000, 4)
    return {
        "conductivity": struct.unpack(">H", data[0:2])[0] * 10,  # µS/cm
        "temperature":  struct.unpack(">H", data[2:4])[0] * 0.1,  # °C
        "tds":          struct.unpack(">H", data[4:6])[0] * 10,  # ppm
        "salinity":     struct.unpack(">H", data[6:8])[0] * 0.1,  # ppt
    }


if __name__ == "__main__":
    with serial.Serial(PORT, 9600, bytesize=8, parity="N", stopbits=1, timeout=1) as port:
        while True:
            print(read_probe(port))
            time.sleep(5)

Arduino example (Uno, Mega, ESP32, ESP8266, MKR, Pico)

The sketch uses the Agrinovo library: in the Arduino IDE open Sketch > Include Library > Manage Libraries and install "Agrinovo". The library picks the serial port for your board, and the pins are listed at the top of AgrinovoDefaultPort.h. DE and RE of the transceiver go to one output pin.

// Agrinovo EC-110 conductivity and salinity probe
// Prints conductivity, temperature, TDS and salinity every 5 seconds.
//
// Sensor cable
//   Red     supply +, 12-24 V DC
//   Black   supply ground, joined to the board's ground
//   Yellow  RS485 A
//   Green   RS485 B
//
// The board talks to the sensor through an RS485 transceiver (MAX485 or
// similar). AgrinovoDefaultPort.h lists the serial pins used on each board.
//
// https://agrinovo.io/products/sensors/aqua/ec/ec-110-probe/

#include <Agrinovo.h>
#include <AgrinovoDefaultPort.h>

const int DE_PIN = 4;  // transceiver DE and RE, tied together

AgrinovoBus bus(AGRINOVO_PORT, DE_PIN);
AgrinovoEC110 probe(bus);  // Modbus address 1

void setup() {
  Serial.begin(115200);
  agrinovoPortBegin(9600);
  bus.begin();
}

void loop() {
  if (probe.read()) {
    Serial.print("Conductivity: ");
    Serial.print(probe.conductivity(), 0);
    Serial.println(" µS/cm");
    Serial.print("Temperature: ");
    Serial.print(probe.temperature(), 1);
    Serial.println(" °C");
    Serial.print("TDS: ");
    Serial.print(probe.tds(), 0);
    Serial.println(" ppm");
    Serial.print("Salinity: ");
    Serial.print(probe.salinity(), 1);
    Serial.println(" ppt");
  } else {
    Serial.println(probe.errorText());
  }
  delay(5000);
}

What the examples read

Both examples send one request, function 0x03, 4 registers from 0x0000, and decode these values:

Value Unit Register Encoding
Conductivity µS/cm 0x0000 Unsigned 16-bit integer, scaled
Temperature °C 0x0001 Unsigned 16-bit integer, scaled
TDS ppm 0x0002 Unsigned 16-bit integer, scaled
Salinity ppt 0x0003 Unsigned 16-bit integer, scaled

Other Modbus tools

The same request works from any Modbus RTU master: pymodbus, minimalmodbus, Node-RED, LabVIEW, a PLC or a third-party data logger with a Modbus port. Use the serial settings, function code and register addresses above.

Logging without writing code

The Omni Genesis controller reads the EC-110 and sends the readings to the Agrinovo cloud over the cellular network, where they can be exported or read through the cloud API.

Datasheet reference

Specifications

Parameter Specification
Conductivity Range 0 to 200,000 µS/cm (0 to 200 mS/cm)
Salinity Range 0 to 100 ppt
TDS Output ppm (composition dependent)
Accuracy ±1.5% FS
Temperature Range 0 to 60°C
Temperature Compensation Automatic
Operating Humidity ≤85% RH
Power Consumption ≤0.5 W
Electrode Constant K = 10.0
Electrode Material Two-graphite
Housing Material PTFE
Output RS485 Modbus-RTU
Supply Voltage 12-24 VDC

Communication Settings

Parameter Value
Protocol Modbus-RTU
Baud Rate 9600 bps
Data Bits 8
Parity None
Stop Bits 1
Default Address 0x01

Register Map

Measurement Registers (Function 0x03)

Address Description Scaling Range
0x0000 Conductivity Value × 10 → µS/cm 0 to 200,000 µS/cm
0x0001 Temperature Value ÷ 10 → °C 0 to 60°C
0x0002 TDS Value × 10 → ppm composition dependent
0x0003 Salinity Value ÷ 10 → ppt 0 to 100 ppt

The salinity-to-TDS ratio depends on water composition. For NaCl-dominant water (seawater, marine aquaculture, industrial brine), the ratio is typically 1.5 to 2.0.

Configuration Registers (Function 0x06)

Address Description Range / Notes
0x0000 Slave Address 0x01 to 0xFE
0x0004 Slope Value × 1000 (default: 1000)

Reading Data

Read 4 registers from 0x0000:

Request:  01 03 00 00 00 04 44 09
Response: 01 03 08 00 1A 00 BF 00 0D 00 02 2B CF

Decoding (tap water example):

Register Hex Decimal Scaling Result
Conductivity 0x001A 26 × 10 260 µS/cm
Temperature 0x00BF 191 ÷ 10 19.1°C
TDS 0x000D 13 × 10 130 ppm
Salinity 0x0002 2 ÷ 10 0.2 ppt

Address Configuration

Change Address (0x01 to 0x02)

Write to register 0x0000:

01 06 00 00 00 02 08 0B

Broadcast Discovery

Use address 0xFE with only one device on the bus.

Calibration

Slope Adjustment

The slope register at 0x0004 trims the sensor against a known reference standard. Slope is stored as value × 1000. To set a slope of 1.200, write 1200 (0x04B0):

01 06 00 04 04 B0 CB 7F

Default slope is 1.000 (raw value 1000).

Revision History

Rev Date Changes
1.0 June 2026 Initial release
1.1 August 2026 Layout revision
1.2 October 2026 New layout

Contact

Agrinovo, https://agrinovo.io/products, [email protected], +972 54 688 8148 (direct / WhatsApp). Specifications subject to change without notice.

Questions about wiring or the protocol?

Tell us which board or logger you are using and what you need to measure.