DO-130 with Arduino, ESP32, Raspberry Pi and Python

How to wire the DO-130 Self-Cleaning Dissolved Oxygen Sensor to your own microcontroller or computer and read it over RS485 Modbus RTU. The tables are the ones in the datasheet, and both code examples are run against the datasheet example frame before this page is published.

Product page Datasheet (PDF) Datasheet (Markdown)

DO-130 Self-Cleaning Dissolved Oxygen Sensor

Connecting the DO-130

The probe is an RS485 device, so it needs a transceiver between it and your board. On a computer or a Raspberry Pi that is a USB to RS485 adapter. On a microcontroller it is a MAX485-type module on a hardware serial port. Power the probe from a supply within the voltage range in the table and join its ground to the ground of the transceiver.

Wire Color Function Description
Red V+ Power Supply (12-24V DC+)
Black GND Power Ground (12-24V DC-)
White RS485-A Data+
Green RS485-B Data-

Note: If A/B labels don’t match your transceiver, swap connections.

Python example (Raspberry Pi, Linux, macOS, Windows)

The script needs only pyserial. It builds the Modbus request itself, checks the CRC of the reply and prints one reading every five seconds. On Windows pass the COM port as the argument.

#!/usr/bin/env python3
"""Read the Agrinovo DO-130 over RS485 Modbus RTU.

Works on a Raspberry Pi or any computer with a USB to RS485 adapter.
Install the serial library once:  pip install pyserial
Run:  python3 do_130.py /dev/ttyUSB0
"""
import struct
import sys
import time

import serial

PORT = sys.argv[1] if len(sys.argv) > 1 else "/dev/ttyUSB0"
ADDRESS = 0x01  # the probe's Modbus address


def crc16(frame: bytes) -> bytes:
    crc = 0xFFFF
    for byte in frame:
        crc ^= byte
        for _ in range(8):
            crc = (crc >> 1) ^ 0xA001 if crc & 1 else crc >> 1
    return struct.pack("<H", crc)


def read_registers(port, address, function, start, count):
    request = struct.pack(">BBHH", address, function, start, count)
    port.reset_input_buffer()
    port.write(request + crc16(request))
    reply = port.read(5 + 2 * count)
    if len(reply) >= 5 and reply[1] == function | 0x80:
        raise IOError(f"probe returned Modbus error {reply[2]}")
    if len(reply) != 5 + 2 * count or crc16(reply[:-2]) != reply[-2:]:
        raise IOError("no valid reply from the probe")
    return reply[3:-2]


def read_probe(port):
    data = read_registers(port, ADDRESS, 0x03, 0x0000, 4)
    return {
        "do_mg_l":       struct.unpack("<f", data[0:4])[0],
        "temperature_c": struct.unpack("<f", data[4:8])[0],
    }


if __name__ == "__main__":
    with serial.Serial(PORT, 9600, bytesize=8, parity="N", stopbits=1, timeout=1) as port:
        while True:
            print(read_probe(port))
            time.sleep(5)

Arduino example (ESP32, Mega and other boards with a second serial port)

The sketch uses no library. On an ESP32 it uses Serial2 on the pins set at the top; on AVR boards it uses Serial1. DE and RE of the transceiver go to one output pin.

// Agrinovo DO-130 over RS485 Modbus RTU.
// For a board with a spare hardware serial port (ESP32, Mega, Leonardo, Due)
// and a MAX485-type transceiver. Power the probe from its own supply, as the
// wiring table says, and join the grounds.

const uint8_t ADDRESS = 0x01;  // the probe's Modbus address
const int DE_PIN = 4;          // transceiver DE and RE tied together
const int RX_PIN = 16;         // ESP32 only
const int TX_PIN = 17;         // ESP32 only

#if defined(ESP32)
#define RS485 Serial2
#else
#define RS485 Serial1
#endif

uint16_t crc16(const uint8_t *frame, size_t length) {
  uint16_t crc = 0xFFFF;
  for (size_t i = 0; i < length; i++) {
    crc ^= frame[i];
    for (int bit = 0; bit < 8; bit++) {
      crc = (crc & 1) ? (crc >> 1) ^ 0xA001 : crc >> 1;
    }
  }
  return crc;
}

// Builds a float from its four bytes, most significant first.
float float32(uint8_t b3, uint8_t b2, uint8_t b1, uint8_t b0) {
  uint32_t bits = ((uint32_t)b3 << 24) | ((uint32_t)b2 << 16) | ((uint32_t)b1 << 8) | b0;
  float value;
  memcpy(&value, &bits, sizeof value);
  return value;
}

// Reads `count` registers into `data`. Returns false when the probe does not
// answer or the checksum is wrong.
bool readRegisters(uint8_t function, uint16_t start, uint16_t count, uint8_t *data) {
  uint8_t request[8] = {ADDRESS, function, (uint8_t)(start >> 8), (uint8_t)start,
                        (uint8_t)(count >> 8), (uint8_t)count, 0, 0};
  uint16_t crc = crc16(request, 6);
  request[6] = crc & 0xFF;
  request[7] = crc >> 8;

  while (RS485.available()) RS485.read();
  digitalWrite(DE_PIN, HIGH);
  RS485.write(request, 8);
  RS485.flush();
  digitalWrite(DE_PIN, LOW);

  uint8_t reply[5 + 2 * 16];
  size_t expected = 5 + 2 * count;
  if (RS485.readBytes(reply, expected) != expected) return false;
  uint16_t received = reply[expected - 2] | ((uint16_t)reply[expected - 1] << 8);
  if (reply[1] != function || crc16(reply, expected - 2) != received) return false;
  memcpy(data, reply + 3, 2 * count);
  return true;
}

void setup() {
  Serial.begin(115200);
  pinMode(DE_PIN, OUTPUT);
  digitalWrite(DE_PIN, LOW);
#if defined(ESP32)
  RS485.begin(9600, SERIAL_8N1, RX_PIN, TX_PIN);
#else
  RS485.begin(9600);
#endif
  RS485.setTimeout(1000);
}

void loop() {
  uint8_t data[2 * 4];
  if (!readRegisters(0x03, 0x0000, 4, data)) {
    Serial.println("no reply");
    delay(5000);
    return;
  }
  Serial.print("do_mg_l=");
  Serial.print(float32(data[3], data[2], data[1], data[0]), 2);
  Serial.print(" temperature_c=");
  Serial.print(float32(data[7], data[6], data[5], data[4]), 2);
  Serial.println();
  delay(5000);
}

What the examples read

Both examples send one request, function 0x03, 4 registers from 0x0000, and decode these values:

Value Unit Register Encoding
do_mg_l mg/L 0x0000 32-bit float, least significant byte first
temperature_c °C 0x0002 32-bit float, least significant byte first

Other Modbus tools

The same request works from any Modbus RTU master: pymodbus, minimalmodbus, Node-RED, LabVIEW, a PLC or a third-party data logger with a Modbus port. Use the serial settings, function code and register addresses above.

Logging without writing code

The Omni Genesis controller reads the DO-130 and sends the readings to the Agrinovo cloud over the cellular network, where they can be exported or read through the cloud API.

Datasheet reference

Specifications

Parameter Specification
DO Unit mg/L
Temperature Unit °C
Salinity Range 1 ppt minimum
Pressure Unit kPa
Supply Voltage 12-24 VDC
Baud Rate Fixed 9600 bps

Communication Settings

Parameter Value
Protocol Modbus-RTU
Baud Rate 9600 bps
Data Bits 8 (LSB first)
Parity None
Stop Bits 1
Checksum CRC16

Register Map

Data Encoding

DO and Temperature are float32 in little-endian byte order.

Measurement Registers (Function 0x03)

Address Registers Description
0x0000 4 DO (float32) + Temp (float32)
0x000A 2 Production Number
0x0010 1 Slave Address
0x0062 1 Salinity (ppt)
0x0064 1 Atmospheric Pressure (kPa × 100)
0x0014 1 Brush Interval (seconds)

Configuration Registers (Function 0x10)

Address Description Notes
0x0010 Set Slave Address Range: 1 to 250
0x0011 Field Calibration Atmospheric pressure × 100 for calibration
0x0013 Manual Brush Trigger Write 0x0000 to clean once
0x0014 Brush Interval Seconds (e.g., 180 = 3 min)
0x0021 Reset Calibration Write 0x0014 to restore factory
0x0062 Set Salinity Value in ppt
0x0064 Set Atmospheric Pressure kPa × 100 (range: 8000-13000)

Reading Data

Read DO + Temperature

Read 4 registers from 0x0000:

Request:  01 03 00 00 00 04 44 09
Response: 01 03 08 [DO_4B] [Temp_4B] [CRC]

Decoding (Little-endian float32):

Example bytes 48 E1 0A 41:

Received Reordered Result
48 E1 0A 41 41 0A E1 48 8.68 mg/L

Address Configuration

Read Address (Broadcast 0xFF)

FF 03 00 10 00 01 90 11

Set New Address

Write to 0x0010 using function 0x10:

[OLD] 10 00 10 00 01 02 [NEW] 00 [CRC]

Salinity Compensation

Read Salinity

01 03 00 62 00 01 25 D4

Response: 0x0A 0x00 = 10 ppt

Write Salinity (10 ppt)

01 10 00 62 00 01 02 0A 00 A8 B2

Calibration

Field Calibration

  1. Place sensor in saturated DO (air) until stable
  2. Write atmospheric pressure to 0x0011

Reset to Factory

Write to 0x0021 to restore factory calibration.

Self-Cleaning Brush

Manual Cleaning Trigger

Trigger immediate brush cleaning cycle by writing 0x0000 to register 0x0013 using function 0x06 (write single register):

Example (Address 0x03):

Request:  03 06 00 13 00 00 79 ED
Response: 03 06 00 13 00 00 79 ED (echo)

The brush cycle typically takes 3-5 seconds to complete.

Set Automatic Cleaning Interval

Write interval (in SECONDS) to register 0x0014 using function 0x06.

Set to 180 seconds (3 minutes) - Address 0x03:

Request:  03 06 00 14 00 B4 C8 5B
Response: 03 06 00 14 00 B4 C8 5B (echo)

Value 0x00B4 = 180 seconds = 3 minutes

Read Current Interval

Read 1 register from 0x0014 using function 0x03:

Example (Address 0x03):

Request:  03 03 00 14 00 01 C5 EC
Response: 03 03 02 00 B4 C1 F3

Response 0x00B4 (big-endian) = 180 seconds

Maintenance Recommendations

Environment Cleaning Interval Seconds Notes
Clean water Every 2-4 hours 7200-14400 Low biofouling
Aquaculture Every 1 hour 3600 Moderate biofouling
High density Every 30 minutes 1800 High biofouling
Wastewater Every 10-15 minutes 600-900 Very high biofouling
Manual override As needed Write 0 to 0x0013 Immediate clean

Calibration Interval and Membrane

  • Recalibration interval depends on the water type and the amount of biofouling and algae. Under normal conditions, calibrate every 6-8 months; the self-cleaning brush handles routine cleaning between calibrations.
  • The membrane cap is replaceable, typically after 2+ years of use.

Revision History

Rev Date Changes
1.0 June 2026 Initial release
1.1 August 2026 Layout revision
1.2 October 2026 New layout; corrected Modbus example CRC values