DO-110 with Arduino, ESP32, Raspberry Pi and Python

How to wire the DO-110 Fluorescent Dissolved Oxygen Sensor (Titanium) to your own microcontroller or computer and read it over RS485 Modbus RTU. The tables are the ones in the datasheet, and both code examples are run against the datasheet example frame before this page is published.

Product page Datasheet (PDF) Datasheet (Markdown) Arduino sketch (.ino) Python script (.py)

DO-110 Fluorescent Dissolved Oxygen Sensor (Titanium)

Connecting the DO-110

The probe is an RS485 device, so it needs a transceiver between it and your board. On a computer or a Raspberry Pi that is a USB to RS485 adapter. On a microcontroller it is a MAX485-type module on a hardware serial port. Power the probe from a supply within the voltage range in the table and join its ground to the ground of the transceiver.

Wire Color Function Description
Black V- Power Ground (12-24V DC-)
Red V+ Power Supply (12-24V DC+)
Yellow/White RS485-A Data+
Green RS485-B Data-

Note: If your RS485 transceiver labels A/B opposite, swap the connections.

Python example (Raspberry Pi, Linux, macOS, Windows)

The script needs only pyserial. It builds the Modbus request itself, checks the CRC of the reply and prints one reading every five seconds. On Windows pass the COM port as the argument.

#!/usr/bin/env python3
"""Read the Agrinovo DO-110 over RS485 Modbus RTU.

Works on a Raspberry Pi or any computer with a USB to RS485 adapter.
Install the serial library once:  pip install pyserial
Run:  python3 do_110.py /dev/ttyUSB0
"""
import struct
import sys
import time

import serial

PORT = sys.argv[1] if len(sys.argv) > 1 else "/dev/ttyUSB0"
ADDRESS = 0x01  # the sensor's Modbus address


def crc16(frame: bytes) -> bytes:
    crc = 0xFFFF
    for byte in frame:
        crc ^= byte
        for _ in range(8):
            crc = (crc >> 1) ^ 0xA001 if crc & 1 else crc >> 1
    return struct.pack("<H", crc)


def read_registers(port, address, function, start, count):
    request = struct.pack(">BBHH", address, function, start, count)
    port.reset_input_buffer()
    port.write(request + crc16(request))
    reply = port.read(5 + 2 * count)
    if len(reply) >= 5 and reply[1] == function | 0x80:
        raise IOError(f"sensor returned Modbus error {reply[2]}")
    if len(reply) != 5 + 2 * count or crc16(reply[:-2]) != reply[-2:]:
        raise IOError("no valid reply from the sensor")
    return reply[3:-2]


def read_probe(port):
    data = read_registers(port, ADDRESS, 0x03, 0x0000, 4)
    return {
        "dissolved_oxygen": struct.unpack("<f", data[0:4])[0],  # mg/L
        "temperature":      struct.unpack("<f", data[4:8])[0],  # °C
    }


if __name__ == "__main__":
    with serial.Serial(PORT, 9600, bytesize=8, parity="N", stopbits=1, timeout=1) as port:
        while True:
            print(read_probe(port))
            time.sleep(5)

Arduino example (Uno, Mega, ESP32, ESP8266, MKR, Pico)

The sketch uses the Agrinovo library: in the Arduino IDE open Sketch > Include Library > Manage Libraries and install "Agrinovo". The library picks the serial port for your board, and the pins are listed at the top of AgrinovoDefaultPort.h. DE and RE of the transceiver go to one output pin.

// Agrinovo DO-110 dissolved oxygen probe
// Prints dissolved oxygen and temperature every 5 seconds.
//
// Sensor cable
//   Red              supply +, 12-24 V DC
//   Black            supply ground, joined to the board's ground
//   Yellow or white  RS485 A
//   Green            RS485 B
//
// The board talks to the sensor through an RS485 transceiver (MAX485 or
// similar). AgrinovoDefaultPort.h lists the serial pins used on each board.
//
// https://agrinovo.io/products/sensors/aqua/do/do-110-titanium-probe/

#include <Agrinovo.h>
#include <AgrinovoDefaultPort.h>

const int DE_PIN = 4;  // transceiver DE and RE, tied together

AgrinovoBus bus(AGRINOVO_PORT, DE_PIN);
AgrinovoDO110 probe(bus);  // Modbus address 1

void setup() {
  Serial.begin(115200);
  agrinovoPortBegin(9600);
  bus.begin();
}

void loop() {
  if (probe.read()) {
    Serial.print("Dissolved oxygen: ");
    Serial.print(probe.dissolvedOxygen(), 2);
    Serial.println(" mg/L");
    Serial.print("Temperature: ");
    Serial.print(probe.temperature(), 2);
    Serial.println(" °C");
  } else {
    Serial.println(probe.errorText());
  }
  delay(5000);
}

What the examples read

Both examples send one request, function 0x03, 4 registers from 0x0000, and decode these values:

Value Unit Register Encoding
Dissolved oxygen mg/L 0x0000 32-bit float, least significant byte first
Temperature °C 0x0002 32-bit float, least significant byte first

Other Modbus tools

The same request works from any Modbus RTU master: pymodbus, minimalmodbus, Node-RED, LabVIEW, a PLC or a third-party data logger with a Modbus port. Use the serial settings, function code and register addresses above.

Logging without writing code

The Omni Genesis controller reads the DO-110 and sends the readings to the Agrinovo cloud over the cellular network, where they can be exported or read through the cloud API.

Datasheet reference

Specifications

Parameter Specification
DO Range 0 to 20 mg/L
Temperature Range 0 to 50°C
Protection Rating IP68
Supply Voltage 12-24 VDC

Compensation

Type Method
Temperature Automatic
Salinity Manual input (ppt)
Pressure Manual input (kPa)

Communication Settings

Parameter Value
Protocol Modbus-RTU
Baud Rate 9600 bps
Data Bits 8
Parity None
Stop Bits 1
Default Address 0x01

Register Map

Data Encoding

DO and Temperature values are float32 in little-endian byte order.

Measurement Registers (Function 0x03)

Address Registers Description
0x0000 4 DO (float32) + Temperature (float32)
0x000A 2 Production Serial Number
0x0010 1 Slave Address (via 0xFF request)
0x0011 2 Field Calibration Parameter
0x0062 1 Salinity (ppt, integer)

Configuration Registers (Function 0x10)

Address Description Notes
0x0010 Set Slave Address Range: 1 to 247
0x0011 Field Calibration Pressure in 0.01 kPa
0x0021 Cancel Field Calibration Write 0x1400
0x0062 Set Salinity Value in ppt

Reading Data

Read DO + Temperature (Float)

Read 4 registers from 0x0000:

Request:  01 03 00 00 00 04 44 09
Response: 01 03 08 48 E1 0A 41 F6 28 B4 41 59 5A

Decoding (Little-endian float32):

Value Bytes Received Reordered Result
DO 48 E1 0A 41 41 0A E1 48 8.68 mg/L
Temperature F6 28 B4 41 41 B4 28 F6 22.52°C

Address Configuration

Read Current Address (Using 0xFF)

Request:  FF 03 00 10 00 01 90 11
Response: FF 03 02 03 00 91 60

Response shows current address = 0x03

Set New Address

Write to 0x0010 using function 0x10:

01 10 00 10 00 01 02 14 00 AB C0

Calibration

Field Calibration (Pressure)

Place probe in air-saturated DO, wait for stability. Write pressure to 0x0011.

Pressure: kPa × 100 (e.g., 101.33 kPa = 10133 = 0x2795)

01 10 00 11 00 01 02 95 27 8A 5B

Cancel Field Calibration

Write 0x1400 to register 0x0021:

01 10 00 21 00 01 02 14 00 AF E1

Maintenance

  • Calibration interval depends on the water type and the amount of biofouling and algae. Under normal conditions, calibrate every 6-8 months.
  • Clean the membrane cap regularly between calibrations.
  • The membrane cap is replaceable, typically after 2+ years of use.

Revision History

Rev Date Changes
1.0 June 2026 Initial release
1.1 August 2026 Layout revision
1.2 October 2026 New layout

Contact

Agrinovo, https://agrinovo.io/products, [email protected], +972 54 688 8148 (direct / WhatsApp). Specifications subject to change without notice.

Questions about wiring or the protocol?

Tell us which board or logger you are using and what you need to measure.