Dissolved oxygen probes come in two families. An electrochemical cell reduces oxygen behind a membrane and measures the current. An optical sensor shines light on a dye and measures how strongly oxygen quenches its glow. Both report mg/L, and nearly every practical difference between them comes from one fact: the first consumes the oxygen it measures and the second does not.
Electrochemical cells: polarographic and galvanic
The classic design is the Clark sensor. A thin gas-permeable membrane is stretched across the tip. Behind it, a film of potassium chloride solution covers a platinum cathode and a silver anode. Oxygen diffuses through the membrane and is reduced at the cathode, while silver at the anode is converted to silver chloride. The steady current that results is proportional to the dissolved oxygen concentration, as described in the amperometry section of Harvey’s Analytical Chemistry 2.1.
A polarographic sensor has the reducing voltage applied by the instrument. A galvanic sensor uses a pair of metals that generate the voltage themselves, so the cell runs whenever oxygen reaches it.
Optical sensors: luminescence quenching
An optical probe has no electrolyte and no electrodes in contact with the sample. A light emitting diode excites an oxygen-sensitive luminescent dye on the sensor cap. The dye releases the energy as light of a longer wavelength, and the luminescence is inversely proportional to the dissolved oxygen concentration. That description is from the EPA Region 4 operating procedure for field measurement of dissolved oxygen. Fluorescence, luminescence and optical are three names used for this method.
Flow dependence
Because a membrane electrode consumes oxygen, it depletes the layer of water against the membrane. The reading falls unless that layer is renewed. The North Carolina DEQ approved procedure for dissolved oxygen puts it as a requirement: for membrane electrodes, movement of water across the membrane is needed for accurate readings, either from flowing water, a stirrer built into the probe, a stir bar, or swirling the probe in the sample.
The same procedure states that movement across the sensor is not required for all luminescence probes and refers the user to the manufacturer’s instructions.
Warm-up
The North Carolina procedure gives the three cases in one paragraph. Polarographic sensors require a 15-minute warm-up. Galvanic sensors require none. Luminescence probes are optical and require none.
Membrane, electrolyte and cap
The anode reaction in an electrochemical cell uses up chloride from the electrolyte and coats the anode. Upkeep therefore means replacing the electrolyte and the membrane on a schedule.
On an optical probe the wearing part is the sensor cap that carries the dye. The EPA procedure directs that an optical probe reading above 0.2 mg/L in a zero-oxygen solution has its cap replaced or receives a two-point calibration.
Drift, fouling and response time
Both families measure oxygen that has to reach a surface, so both are affected by what grows on that surface. Cleaning is part of the routine for either type, and our datasheets tie the calibration interval to the water type and the amount of biofouling and algae.
In either family the response is limited by how fast oxygen diffuses to the sensing element. Use the t90 on the datasheet of the probe you are considering instead of a figure for the family.
Drift is quantified the same way for both families. Read the probe at 100% saturation after a deployment, before adjusting anything, and compare the reading with the solubility table value.
Salinity and pressure compensation
A dissolved oxygen sensor of either type responds to the partial pressure of oxygen. Turning that into mg/L needs three corrections, because the solubility of oxygen falls as temperature rises, falls as salinity rises and falls as atmospheric pressure drops (EPA Region 4 procedure, section 4.1).
- Temperature is measured by the probe.
- Salinity has to be supplied, from a conductivity sensor or as a value you enter. The EPA procedure requires the final reading to be corrected for salinity in saline water.
- Barometric pressure has to be the true local pressure. Weather services report pressure corrected to sea level, and the North Carolina procedure converts it with true pressure = corrected pressure minus 2.5 mmHg for every 100 ft of altitude.
In the North Carolina table, water at 20 °C holds 9.09 mg/L at 760 mmHg. At 676 mmHg, about 3,190 ft, the correction factor is 0.89, so the same water holds about 8.1 mg/L.
Calibration in water-saturated air
Air that is saturated with water vapor holds oxygen at the same partial pressure as air-saturated water, so it serves as a 100% saturation standard without a bubbler.
- Place the probe in a 100% relative humidity environment open to ambient pressure, such as a calibration cup or bag with a wet sponge. Dry air gives errant readings.
- Allow 10 to 15 minutes for the temperature and reading to stabilize.
- Remove water droplets from the sensing face.
- Enter the true barometric pressure and store the point.
- Verify. The EPA procedure accepts a post-calibration reading within ±0.2 mg/L of the solubility table value for that temperature and pressure.
When readings below 1 mg/L are expected, add a zero check. Dissolve 1 gram of sodium sulfite in 1 liter of deionized water, immerse the probe and confirm it reads no more than 0.2 mg/L. Rinse thoroughly afterwards.
Which principle each Agrinovo DO probe uses
All four probes are optical. The DO-100 datasheet names the method as fluorescence, and the DO-110, DO-130 and DO-P100 are specified as fluorescent dissolved oxygen sensors. All four have RS485 Modbus RTU output and a 12 to 24 V DC supply.
| DO-100 | DO-110 | DO-130 | DO-P100 | |
|---|---|---|---|---|
| Principle | Fluorescence | Fluorescence | Fluorescence | Fluorescence |
| Build | 316L stainless steel body | Titanium alloy body | Self-cleaning brush with configurable interval | PE body |
| Compensation | Temperature automatic, pressure by manual input (default 101.3 kPa), salinity input in PSU | Temperature automatic, salinity in ppt, pressure in kPa | Salinity in ppt, atmospheric pressure 80 to 130 kPa | Temperature automatic (PT1000) |
| Calibration | Zero point in zero-oxygen water and slope point in air-saturated conditions, one command each | Field calibration in air-saturated conditions by writing the barometric pressure | Field calibration in air-saturated conditions by writing the atmospheric pressure, with a command to restore factory calibration | One-point calibration |
| Membrane cap | Replaceable, typically after 2 or more years | Replaceable, typically after 2 or more years | Replaceable, typically after 2 or more years | Non-replaceable |
The measurement figures stated for each:
- DO-100. 0 to 20 mg/L and 0 to 200% saturation, accuracy ±2% of range, repeatability ±1%, t90 of about 60 seconds, 0 to 50 °C, IP68.
- DO-110. 0 to 20 mg/L, 0 to 50 °C, IP68, with accuracy specified as ±2% F.S. and response under 60 seconds. This is the probe for seawater.
- DO-P100. 0 to 20 mg/L and 0 to 200% saturation, accuracy ±2% F.S., response under 60 seconds, 0 to 40 °C, IP68. It is built for freshwater.
- DO-130. Reports mg/L and °C. The brush cycle takes 3 to 5 seconds, and the datasheet recommends an interval of 2 to 4 hours in clean water, 1 hour in aquaculture, 30 minutes in high-density culture and 10 to 15 minutes in wastewater.
The DO-100, DO-110 and DO-130 datasheets give the same calibration guidance: the interval depends on the water type and the amount of biofouling and algae, and under normal conditions it is every 6 to 8 months. On the DO-110 and DO-130, calibration is one write of the true barometric pressure in kPa with the probe sitting in water-saturated air.
Register maps and example frames are on the developer pages for the DO-100 and DO-130.
Choosing between them
An optical probe fits still or slow water, intermittent power and long unattended deployment. A membrane electrode remains a sound laboratory instrument where the sample is stirred and the cell is serviced on a schedule.
Housing material, species oxygen requirements and probe placement in ponds, tanks and cages are covered in how to choose a dissolved oxygen sensor for fish farming. For what to do with the readings at night, see preventing overnight fish kills.