ORP-100 with Arduino, ESP32, Raspberry Pi and Python
How to wire the ORP-100 Digital ORP Probe - RS485, Auto Temp Comp to your own microcontroller or computer and read it over RS485 Modbus RTU. The tables are the ones in the datasheet, and both code examples are run against the datasheet example frame before this page is published.
Product page Datasheet (PDF) Datasheet (Markdown) Arduino sketch (.ino) Python script (.py)
Connecting the ORP-100
The probe is an RS485 device, so it needs a transceiver between it and your board. On a computer or a Raspberry Pi that is a USB to RS485 adapter. On a microcontroller it is a MAX485-type module on a hardware serial port. Power the probe from a supply within the voltage range in the table and join its ground to the ground of the transceiver.
| Wire Color | Function | Description |
|---|---|---|
| Red | V+ | Power Supply (12-24V DC) |
| Black | GND | Power Ground |
| Yellow | RS485-A | Data+ |
| Green | RS485-B | Data- |
Python example (Raspberry Pi, Linux, macOS, Windows)
The script needs only pyserial. It builds the Modbus request itself, checks the CRC of the reply and prints one reading every five seconds. On Windows pass the COM port as the argument.
#!/usr/bin/env python3
"""Read the Agrinovo ORP-100 over RS485 Modbus RTU.
Works on a Raspberry Pi or any computer with a USB to RS485 adapter.
Install the serial library once: pip install pyserial
Run: python3 orp_100.py /dev/ttyUSB0
"""
import struct
import sys
import time
import serial
PORT = sys.argv[1] if len(sys.argv) > 1 else "/dev/ttyUSB0"
ADDRESS = 0x01 # the sensor's Modbus address
def crc16(frame: bytes) -> bytes:
crc = 0xFFFF
for byte in frame:
crc ^= byte
for _ in range(8):
crc = (crc >> 1) ^ 0xA001 if crc & 1 else crc >> 1
return struct.pack("<H", crc)
def read_registers(port, address, function, start, count):
request = struct.pack(">BBHH", address, function, start, count)
port.reset_input_buffer()
port.write(request + crc16(request))
reply = port.read(5 + 2 * count)
if len(reply) >= 5 and reply[1] == function | 0x80:
raise IOError(f"sensor returned Modbus error {reply[2]}")
if len(reply) != 5 + 2 * count or crc16(reply[:-2]) != reply[-2:]:
raise IOError("no valid reply from the sensor")
return reply[3:-2]
def read_probe(port):
data = read_registers(port, ADDRESS, 0x03, 0x0000, 1)
return {
"orp": struct.unpack(">h", data[0:2])[0], # mV
}
if __name__ == "__main__":
with serial.Serial(PORT, 9600, bytesize=8, parity="N", stopbits=1, timeout=1) as port:
while True:
print(read_probe(port))
time.sleep(5) Arduino example (Uno, Mega, ESP32, ESP8266, MKR, Pico)
The sketch uses the Agrinovo library: in the Arduino IDE open Sketch > Include Library > Manage Libraries and install "Agrinovo". The library picks the serial port for your board, and the pins are listed at the top of AgrinovoDefaultPort.h. DE and RE of the transceiver go to one output pin.
// Agrinovo ORP-100 ORP probe
// Prints ORP every 5 seconds.
//
// Sensor cable
// Red supply +, 12-24 V DC
// Black supply ground, joined to the board's ground
// Yellow RS485 A
// Green RS485 B
//
// The board talks to the sensor through an RS485 transceiver (MAX485 or
// similar). AgrinovoDefaultPort.h lists the serial pins used on each board.
//
// https://agrinovo.io/products/sensors/aqua/orp/orp-100-probe/
#include <Agrinovo.h>
#include <AgrinovoDefaultPort.h>
const int DE_PIN = 4; // transceiver DE and RE, tied together
AgrinovoBus bus(AGRINOVO_PORT, DE_PIN);
AgrinovoORP100 probe(bus); // Modbus address 1
void setup() {
Serial.begin(115200);
agrinovoPortBegin(9600);
bus.begin();
}
void loop() {
if (probe.read()) {
Serial.print("ORP: ");
Serial.print(probe.orp(), 0);
Serial.println(" mV");
} else {
Serial.println(probe.errorText());
}
delay(5000);
} What the examples read
Both examples send one request, function 0x03, 1 registers from 0x0000, and decode these values:
| Value | Unit | Register | Encoding |
|---|---|---|---|
| ORP | mV | 0x0000 | Signed 16-bit integer, scaled |
Other Modbus tools
The same request works from any Modbus RTU master: pymodbus, minimalmodbus, Node-RED, LabVIEW, a PLC or a third-party data logger with a Modbus port. Use the serial settings, function code and register addresses above.
Logging without writing code
The Omni Genesis controller reads the ORP-100 and sends the readings to the Agrinovo cloud over the cellular network, where they can be exported or read through the cloud API.
Datasheet reference
Specifications
| Parameter | Specification |
|---|---|
| Measurement Range | -1999 to +1999 mV |
| Resolution | 1 mV |
| Accuracy | ±3 mV |
| Stability | ≤2 mV / 24 hours |
| Power Consumption | ≤0.5W |
| Supply Voltage | 12-24 VDC |
| Operating Temp | 0 to 60°C |
| Operating Humidity | ≤85% RH |
| Protection | IP68 |
| Dimensions | Ø28 mm × 160 mm |
| Mounting Thread | 3/4 inch NPT, front and rear |
Communication Settings
| Parameter | Value |
|---|---|
| Protocol | Modbus-RTU |
| Baud Rate | 9600 bps |
| Data Bits | 8 |
| Parity | None |
| Stop Bits | 1 |
| Default Address | 0x01 |
Register Map
Measurement Registers (Function 0x03)
| Address | Description | Scaling |
|---|---|---|
| 0x0000 | ORP Value (mV) | Direct (signed 16-bit) |
Configuration Registers (Function 0x06)
| Address | Description | Notes |
|---|---|---|
| 0x000A | Calibration Offset | Signed offset in mV |
| 0x0050 | Slave Address | 0x01 to 0xFE |
Reading ORP Value
Read 1 register from 0x0000:
Request: 01 03 00 00 00 01 84 0A
Response: 01 03 02 02 AE 38 98
Decoding:
| Hex | Decimal | Result |
|---|---|---|
| 0x02AE | 686 | 686 mV |
Negative values use two’s complement. For example, 0xFF9C = -100 mV.
Single-Point Calibration
ORP uses single-point offset calibration via register 0x000A.
Procedure
- Prepare an ORP standard solution (e.g. +220 mV)
- Clean the electrode with distilled water
- Immerse in the standard solution and wait for stabilization
- Read the current sensor value
- Calculate the offset:
offset = standard - reading - Write the offset to register 0x000A
Example
If the standard is +220 mV and the sensor reads +200 mV, the offset is +20:
01 06 00 0A 00 14 A9 C7
If the sensor reads +240 mV, the offset is -20 (0xFFEC in two’s complement):
01 06 00 0A FF EC E9 B5
Address Configuration
Change Address (0x01 → 0x02)
01 06 00 50 00 02 08 1A
Broadcast Discovery
Use address 0xFE with only one device connected.
Revision History
| Rev | Date | Changes |
|---|---|---|
| 1.0 | June 2026 | Initial release |
| 1.1 | August 2026 | Two-page layout; CRC values added to calibration and address examples |
| 1.2 | October 2026 | New layout |
Contact
Agrinovo, https://agrinovo.io/products, [email protected], +972 54 688 8148 (direct / WhatsApp). Specifications subject to change without notice.
Questions about wiring or the protocol?
Tell us which board or logger you are using and what you need to measure.