CO2-100 with Arduino, ESP32, Raspberry Pi and Python

How to wire the CO2-100 Dissolved CO2 Sensor - 0.2-2,200 ppm, RS485 to your own microcontroller or computer and read it over RS485 Modbus RTU. The tables are the ones in the datasheet, and both code examples are run against the datasheet example frame before this page is published.

Product page Datasheet (PDF) Datasheet (Markdown)

CO2-100 Dissolved CO2 Sensor - 0.2-2,200 ppm, RS485

Connecting the CO2-100

The probe is an RS485 device, so it needs a transceiver between it and your board. On a computer or a Raspberry Pi that is a USB to RS485 adapter. On a microcontroller it is a MAX485-type module on a hardware serial port. Power the probe from a supply within the voltage range in the table and join its ground to the ground of the transceiver.

Color Function Description
Red V+ Power Supply (9-27V DC)
Black GND Power Ground
Yellow RS485-A Data+ (T/R+)
White RS485-B Data- (T/R-)

Python example (Raspberry Pi, Linux, macOS, Windows)

The script needs only pyserial. It builds the Modbus request itself, checks the CRC of the reply and prints one reading every five seconds. On Windows pass the COM port as the argument.

#!/usr/bin/env python3
"""Read the Agrinovo CO2-100 over RS485 Modbus RTU.

Works on a Raspberry Pi or any computer with a USB to RS485 adapter.
Install the serial library once:  pip install pyserial
Run:  python3 co2_100.py /dev/ttyUSB0
"""
import struct
import sys
import time

import serial

PORT = sys.argv[1] if len(sys.argv) > 1 else "/dev/ttyUSB0"
ADDRESS = 0x01  # the probe's Modbus address


def crc16(frame: bytes) -> bytes:
    crc = 0xFFFF
    for byte in frame:
        crc ^= byte
        for _ in range(8):
            crc = (crc >> 1) ^ 0xA001 if crc & 1 else crc >> 1
    return struct.pack("<H", crc)


def read_registers(port, address, function, start, count):
    request = struct.pack(">BBHH", address, function, start, count)
    port.reset_input_buffer()
    port.write(request + crc16(request))
    reply = port.read(5 + 2 * count)
    if len(reply) >= 5 and reply[1] == function | 0x80:
        raise IOError(f"probe returned Modbus error {reply[2]}")
    if len(reply) != 5 + 2 * count or crc16(reply[:-2]) != reply[-2:]:
        raise IOError("no valid reply from the probe")
    return reply[3:-2]


def read_probe(port):
    data = read_registers(port, ADDRESS, 0x03, 0x0000, 10)
    return {
        "co2_ppm":       struct.unpack(">f", data[2:4] + data[0:2])[0],
        "electrode_mv":  struct.unpack(">f", data[6:8] + data[4:6])[0],
        "temperature_c": struct.unpack(">f", data[18:20] + data[16:18])[0],
    }


if __name__ == "__main__":
    with serial.Serial(PORT, 9600, bytesize=8, parity="N", stopbits=1, timeout=1) as port:
        while True:
            print(read_probe(port))
            time.sleep(5)

Arduino example (ESP32, Mega and other boards with a second serial port)

The sketch uses no library. On an ESP32 it uses Serial2 on the pins set at the top; on AVR boards it uses Serial1. DE and RE of the transceiver go to one output pin.

// Agrinovo CO2-100 over RS485 Modbus RTU.
// For a board with a spare hardware serial port (ESP32, Mega, Leonardo, Due)
// and a MAX485-type transceiver. Power the probe from its own supply, as the
// wiring table says, and join the grounds.

const uint8_t ADDRESS = 0x01;  // the probe's Modbus address
const int DE_PIN = 4;          // transceiver DE and RE tied together
const int RX_PIN = 16;         // ESP32 only
const int TX_PIN = 17;         // ESP32 only

#if defined(ESP32)
#define RS485 Serial2
#else
#define RS485 Serial1
#endif

uint16_t crc16(const uint8_t *frame, size_t length) {
  uint16_t crc = 0xFFFF;
  for (size_t i = 0; i < length; i++) {
    crc ^= frame[i];
    for (int bit = 0; bit < 8; bit++) {
      crc = (crc & 1) ? (crc >> 1) ^ 0xA001 : crc >> 1;
    }
  }
  return crc;
}

// Builds a float from its four bytes, most significant first.
float float32(uint8_t b3, uint8_t b2, uint8_t b1, uint8_t b0) {
  uint32_t bits = ((uint32_t)b3 << 24) | ((uint32_t)b2 << 16) | ((uint32_t)b1 << 8) | b0;
  float value;
  memcpy(&value, &bits, sizeof value);
  return value;
}

// Reads `count` registers into `data`. Returns false when the probe does not
// answer or the checksum is wrong.
bool readRegisters(uint8_t function, uint16_t start, uint16_t count, uint8_t *data) {
  uint8_t request[8] = {ADDRESS, function, (uint8_t)(start >> 8), (uint8_t)start,
                        (uint8_t)(count >> 8), (uint8_t)count, 0, 0};
  uint16_t crc = crc16(request, 6);
  request[6] = crc & 0xFF;
  request[7] = crc >> 8;

  while (RS485.available()) RS485.read();
  digitalWrite(DE_PIN, HIGH);
  RS485.write(request, 8);
  RS485.flush();
  digitalWrite(DE_PIN, LOW);

  uint8_t reply[5 + 2 * 16];
  size_t expected = 5 + 2 * count;
  if (RS485.readBytes(reply, expected) != expected) return false;
  uint16_t received = reply[expected - 2] | ((uint16_t)reply[expected - 1] << 8);
  if (reply[1] != function || crc16(reply, expected - 2) != received) return false;
  memcpy(data, reply + 3, 2 * count);
  return true;
}

void setup() {
  Serial.begin(115200);
  pinMode(DE_PIN, OUTPUT);
  digitalWrite(DE_PIN, LOW);
#if defined(ESP32)
  RS485.begin(9600, SERIAL_8N1, RX_PIN, TX_PIN);
#else
  RS485.begin(9600);
#endif
  RS485.setTimeout(1000);
}

void loop() {
  uint8_t data[2 * 10];
  if (!readRegisters(0x03, 0x0000, 10, data)) {
    Serial.println("no reply");
    delay(5000);
    return;
  }
  Serial.print("co2_ppm=");
  Serial.print(float32(data[2], data[3], data[0], data[1]), 2);
  Serial.print(" electrode_mv=");
  Serial.print(float32(data[6], data[7], data[4], data[5]), 1);
  Serial.print(" temperature_c=");
  Serial.print(float32(data[18], data[19], data[16], data[17]), 1);
  Serial.println();
  delay(5000);
}

What the examples read

Both examples send one request, function 0x03, 10 registers from 0x0000, and decode these values:

Value Unit Register Encoding
co2_ppm ppm 0x0000 32-bit float in two registers, low word first
electrode_mv mV 0x0002 32-bit float in two registers, low word first
temperature_c °C 0x0008 32-bit float in two registers, low word first

Other Modbus tools

The same request works from any Modbus RTU master: pymodbus, minimalmodbus, Node-RED, LabVIEW, a PLC or a third-party data logger with a Modbus port. Use the serial settings, function code and register addresses above.

Logging without writing code

The Omni Genesis controller reads the CO2-100 and sends the readings to the Agrinovo cloud over the cellular network, where they can be exported or read through the cloud API.

Datasheet reference

Specifications

Parameter Specification
CO₂ Range 0.2-2200 ppm (configurable to 20000 ppm)
Resolution 0.01 ppm
Accuracy ±5% FS
Temperature Range -10 to 110°C
Temperature Accuracy ±0.5°C
Temperature Compensation Automatic (NTC10K) or manual
Working pH Range 4.8-5.2
Supply Voltage 9-27 VDC
Power Consumption ≤0.5W
Isolation 2500 Vrms
Housing PPS, NPT 3/4” front and back thread
Dimensions Ø30 mm, 220 mm length

Communication Settings

Parameter Value
Protocol Modbus-RTU
Baud Rate 9600 bps (default)
Data Bits 8
Parity None
Stop Bits 1
Default Address 0x01

Data Format

  • Type: 32-bit Single Precision Float (IEEE-754)
  • Byte Order: Word-swapped, big-endian within each word

Each measurement spans two registers. Assemble the float from the second register (high word) followed by the first register (low word). Supported baud rates: 1200, 2400, 4800, 9600, 19200.

Register Map

Measurement Registers (Function 0x03, float32)

Address (Hex) Description Range Unit
0x0000-0x0001 CO₂ Concentration 0-20000 ppm
0x0002-0x0003 Electrode Signal -750 to 750 mV
0x0008-0x0009 Temperature -10 to 110 °C

Parameter Registers (Function 0x03 read / 0x06 write)

Address (Hex) Description Range / Values Default
0x0019 Calibration Status Bitfield (read) -
0x001C Electrode Efficiency 70.0-130.0% (÷10) 100.0%
0x001D Calibrated Points 0-5 0
0x001E Device Address 1-247 (255 = broadcast) 1
0x001F Baud Rate 0:1200 1:2400 2:4800 3:9600 4:19200 3 (9600)
0x0020 Temp Comp Type 0: Manual, 1: Automatic 1
0x0021 Temp Setpoint / Offset -10.0 to 110.0°C (×10) 25.0°C / 0.0°C
0x0023 Ion Valence 0: Unset, 1: Monovalent, 2: Divalent -

Information Registers (Function 0x03 read)

Address (Hex) Description Notes
0x0040 Work Mode 0x0010 Measure, 0x0050 Setup, 0x0060 Calibration
0x0043 Calibration Status / Operation Read status, write standard code
0x0045 Model Number 0x1210
0x0046 Software Version BCD
0x0047 Hardware Version BCD
0x0048-0x0049 Serial Number BCD

Reading Data

Read CO₂ + Electrode Signal + Temperature

Read 10 registers from 0x0000:

Request:  01 03 00 00 00 0A C5 CD
Response: 01 03 14 00 00 41 20 33 33 42 C8 00 00 00 00
          00 00 00 00 E3 E8 41 C7 43 0C

Decoding (word-swapped float32):

Value Registers Reordered Result
CO₂ 00 00 / 41 20 41 20 00 00 10.00 ppm
Electrode Signal 33 33 / 42 C8 42 C8 33 33 100.1 mV
Temperature E3 E8 / 41 C7 41 C7 E3 E8 25.0°C

Address Configuration

Change Address (0x01 → 0x02)

Write to register 0x001E:

01 06 00 1E 00 02 68 0D

Note: Communication will be lost until reconnecting with the new address.

Broadcast Discovery

Use address 255 (0xFF) with only one device connected.

Calibration

The sensor supports up to 5 calibration points. Calibration is performed over RS485 by placing the electrode in a standard solution and writing the matching standard code to the calibration register (0x0043).

Standard Solution Codes

Standard Code
0.1 ppm 0x0002
1 ppm 0x0004
10 ppm 0x0008
100 ppm 0x0010
1000 ppm 0x0020

Calibration Procedure

  1. Place the electrode in the standard solution.
  2. Write the standard code to register 0x0043. Example (1 ppm point):
01 06 00 43 00 04 79 DD
  1. Poll register 0x0043 to read calibration status:
Value Meaning
0 Success (returned to measurement mode)
1 In progress (wait and re-read)
2 Incorrect standard solution value
3 Signal unstable or out of range after 180s
4 Sensor performance out of allowed range
  1. Repeat for additional points.

Clear Calibration

Write 0x7FFF to register 0x0043 to clear all calibration data.

Factory Reset

Enter setup mode, then restore defaults:

01 06 00 40 00 50 88 22   (enter setup mode)
01 06 00 41 7F FF B9 AE   (restore defaults and restart)

Revision History

Rev Date Changes
1.0 June 2026 Initial release
1.1 August 2026 Layout revision
1.2 October 2026 New layout; corrected Modbus example CRC values