CO2-100 with Arduino, ESP32, Raspberry Pi and Python
How to wire the CO2-100 Dissolved CO2 Sensor - 0.2-2,200 ppm, RS485 to your own microcontroller or computer and read it over RS485 Modbus RTU. The tables are the ones in the datasheet, and both code examples are run against the datasheet example frame before this page is published.
Connecting the CO2-100
The probe is an RS485 device, so it needs a transceiver between it and your board. On a computer or a Raspberry Pi that is a USB to RS485 adapter. On a microcontroller it is a MAX485-type module on a hardware serial port. Power the probe from a supply within the voltage range in the table and join its ground to the ground of the transceiver.
| Color | Function | Description |
|---|---|---|
| Red | V+ | Power Supply (9-27V DC) |
| Black | GND | Power Ground |
| Yellow | RS485-A | Data+ (T/R+) |
| White | RS485-B | Data- (T/R-) |
Python example (Raspberry Pi, Linux, macOS, Windows)
The script needs only pyserial. It builds the Modbus request itself, checks the CRC of the reply and prints one reading every five seconds. On Windows pass the COM port as the argument.
#!/usr/bin/env python3
"""Read the Agrinovo CO2-100 over RS485 Modbus RTU.
Works on a Raspberry Pi or any computer with a USB to RS485 adapter.
Install the serial library once: pip install pyserial
Run: python3 co2_100.py /dev/ttyUSB0
"""
import struct
import sys
import time
import serial
PORT = sys.argv[1] if len(sys.argv) > 1 else "/dev/ttyUSB0"
ADDRESS = 0x01 # the probe's Modbus address
def crc16(frame: bytes) -> bytes:
crc = 0xFFFF
for byte in frame:
crc ^= byte
for _ in range(8):
crc = (crc >> 1) ^ 0xA001 if crc & 1 else crc >> 1
return struct.pack("<H", crc)
def read_registers(port, address, function, start, count):
request = struct.pack(">BBHH", address, function, start, count)
port.reset_input_buffer()
port.write(request + crc16(request))
reply = port.read(5 + 2 * count)
if len(reply) >= 5 and reply[1] == function | 0x80:
raise IOError(f"probe returned Modbus error {reply[2]}")
if len(reply) != 5 + 2 * count or crc16(reply[:-2]) != reply[-2:]:
raise IOError("no valid reply from the probe")
return reply[3:-2]
def read_probe(port):
data = read_registers(port, ADDRESS, 0x03, 0x0000, 10)
return {
"co2_ppm": struct.unpack(">f", data[2:4] + data[0:2])[0],
"electrode_mv": struct.unpack(">f", data[6:8] + data[4:6])[0],
"temperature_c": struct.unpack(">f", data[18:20] + data[16:18])[0],
}
if __name__ == "__main__":
with serial.Serial(PORT, 9600, bytesize=8, parity="N", stopbits=1, timeout=1) as port:
while True:
print(read_probe(port))
time.sleep(5) Arduino example (ESP32, Mega and other boards with a second serial port)
The sketch uses no library. On an ESP32 it uses Serial2 on the pins set at the top; on AVR boards it uses Serial1. DE and RE of the transceiver go to one output pin.
// Agrinovo CO2-100 over RS485 Modbus RTU.
// For a board with a spare hardware serial port (ESP32, Mega, Leonardo, Due)
// and a MAX485-type transceiver. Power the probe from its own supply, as the
// wiring table says, and join the grounds.
const uint8_t ADDRESS = 0x01; // the probe's Modbus address
const int DE_PIN = 4; // transceiver DE and RE tied together
const int RX_PIN = 16; // ESP32 only
const int TX_PIN = 17; // ESP32 only
#if defined(ESP32)
#define RS485 Serial2
#else
#define RS485 Serial1
#endif
uint16_t crc16(const uint8_t *frame, size_t length) {
uint16_t crc = 0xFFFF;
for (size_t i = 0; i < length; i++) {
crc ^= frame[i];
for (int bit = 0; bit < 8; bit++) {
crc = (crc & 1) ? (crc >> 1) ^ 0xA001 : crc >> 1;
}
}
return crc;
}
// Builds a float from its four bytes, most significant first.
float float32(uint8_t b3, uint8_t b2, uint8_t b1, uint8_t b0) {
uint32_t bits = ((uint32_t)b3 << 24) | ((uint32_t)b2 << 16) | ((uint32_t)b1 << 8) | b0;
float value;
memcpy(&value, &bits, sizeof value);
return value;
}
// Reads `count` registers into `data`. Returns false when the probe does not
// answer or the checksum is wrong.
bool readRegisters(uint8_t function, uint16_t start, uint16_t count, uint8_t *data) {
uint8_t request[8] = {ADDRESS, function, (uint8_t)(start >> 8), (uint8_t)start,
(uint8_t)(count >> 8), (uint8_t)count, 0, 0};
uint16_t crc = crc16(request, 6);
request[6] = crc & 0xFF;
request[7] = crc >> 8;
while (RS485.available()) RS485.read();
digitalWrite(DE_PIN, HIGH);
RS485.write(request, 8);
RS485.flush();
digitalWrite(DE_PIN, LOW);
uint8_t reply[5 + 2 * 16];
size_t expected = 5 + 2 * count;
if (RS485.readBytes(reply, expected) != expected) return false;
uint16_t received = reply[expected - 2] | ((uint16_t)reply[expected - 1] << 8);
if (reply[1] != function || crc16(reply, expected - 2) != received) return false;
memcpy(data, reply + 3, 2 * count);
return true;
}
void setup() {
Serial.begin(115200);
pinMode(DE_PIN, OUTPUT);
digitalWrite(DE_PIN, LOW);
#if defined(ESP32)
RS485.begin(9600, SERIAL_8N1, RX_PIN, TX_PIN);
#else
RS485.begin(9600);
#endif
RS485.setTimeout(1000);
}
void loop() {
uint8_t data[2 * 10];
if (!readRegisters(0x03, 0x0000, 10, data)) {
Serial.println("no reply");
delay(5000);
return;
}
Serial.print("co2_ppm=");
Serial.print(float32(data[2], data[3], data[0], data[1]), 2);
Serial.print(" electrode_mv=");
Serial.print(float32(data[6], data[7], data[4], data[5]), 1);
Serial.print(" temperature_c=");
Serial.print(float32(data[18], data[19], data[16], data[17]), 1);
Serial.println();
delay(5000);
} What the examples read
Both examples send one request, function 0x03, 10 registers from 0x0000, and decode these values:
| Value | Unit | Register | Encoding |
|---|---|---|---|
co2_ppm | ppm | 0x0000 | 32-bit float in two registers, low word first |
electrode_mv | mV | 0x0002 | 32-bit float in two registers, low word first |
temperature_c | °C | 0x0008 | 32-bit float in two registers, low word first |
Other Modbus tools
The same request works from any Modbus RTU master: pymodbus, minimalmodbus, Node-RED, LabVIEW, a PLC or a third-party data logger with a Modbus port. Use the serial settings, function code and register addresses above.
Logging without writing code
The Omni Genesis controller reads the CO2-100 and sends the readings to the Agrinovo cloud over the cellular network, where they can be exported or read through the cloud API.
Datasheet reference
Specifications
| Parameter | Specification |
|---|---|
| CO₂ Range | 0.2-2200 ppm (configurable to 20000 ppm) |
| Resolution | 0.01 ppm |
| Accuracy | ±5% FS |
| Temperature Range | -10 to 110°C |
| Temperature Accuracy | ±0.5°C |
| Temperature Compensation | Automatic (NTC10K) or manual |
| Working pH Range | 4.8-5.2 |
| Supply Voltage | 9-27 VDC |
| Power Consumption | ≤0.5W |
| Isolation | 2500 Vrms |
| Housing | PPS, NPT 3/4” front and back thread |
| Dimensions | Ø30 mm, 220 mm length |
Communication Settings
| Parameter | Value |
|---|---|
| Protocol | Modbus-RTU |
| Baud Rate | 9600 bps (default) |
| Data Bits | 8 |
| Parity | None |
| Stop Bits | 1 |
| Default Address | 0x01 |
Data Format
- Type: 32-bit Single Precision Float (IEEE-754)
- Byte Order: Word-swapped, big-endian within each word
Each measurement spans two registers. Assemble the float from the second register (high word) followed by the first register (low word). Supported baud rates: 1200, 2400, 4800, 9600, 19200.
Register Map
Measurement Registers (Function 0x03, float32)
| Address (Hex) | Description | Range | Unit |
|---|---|---|---|
| 0x0000-0x0001 | CO₂ Concentration | 0-20000 | ppm |
| 0x0002-0x0003 | Electrode Signal | -750 to 750 | mV |
| 0x0008-0x0009 | Temperature | -10 to 110 | °C |
Parameter Registers (Function 0x03 read / 0x06 write)
| Address (Hex) | Description | Range / Values | Default |
|---|---|---|---|
| 0x0019 | Calibration Status | Bitfield (read) | - |
| 0x001C | Electrode Efficiency | 70.0-130.0% (÷10) | 100.0% |
| 0x001D | Calibrated Points | 0-5 | 0 |
| 0x001E | Device Address | 1-247 (255 = broadcast) | 1 |
| 0x001F | Baud Rate | 0:1200 1:2400 2:4800 3:9600 4:19200 | 3 (9600) |
| 0x0020 | Temp Comp Type | 0: Manual, 1: Automatic | 1 |
| 0x0021 | Temp Setpoint / Offset | -10.0 to 110.0°C (×10) | 25.0°C / 0.0°C |
| 0x0023 | Ion Valence | 0: Unset, 1: Monovalent, 2: Divalent | - |
Information Registers (Function 0x03 read)
| Address (Hex) | Description | Notes |
|---|---|---|
| 0x0040 | Work Mode | 0x0010 Measure, 0x0050 Setup, 0x0060 Calibration |
| 0x0043 | Calibration Status / Operation | Read status, write standard code |
| 0x0045 | Model Number | 0x1210 |
| 0x0046 | Software Version | BCD |
| 0x0047 | Hardware Version | BCD |
| 0x0048-0x0049 | Serial Number | BCD |
Reading Data
Read CO₂ + Electrode Signal + Temperature
Read 10 registers from 0x0000:
Request: 01 03 00 00 00 0A C5 CD
Response: 01 03 14 00 00 41 20 33 33 42 C8 00 00 00 00
00 00 00 00 E3 E8 41 C7 43 0C
Decoding (word-swapped float32):
| Value | Registers | Reordered | Result |
|---|---|---|---|
| CO₂ | 00 00 / 41 20 | 41 20 00 00 | 10.00 ppm |
| Electrode Signal | 33 33 / 42 C8 | 42 C8 33 33 | 100.1 mV |
| Temperature | E3 E8 / 41 C7 | 41 C7 E3 E8 | 25.0°C |
Address Configuration
Change Address (0x01 → 0x02)
Write to register 0x001E:
01 06 00 1E 00 02 68 0D
Note: Communication will be lost until reconnecting with the new address.
Broadcast Discovery
Use address 255 (0xFF) with only one device connected.
Calibration
The sensor supports up to 5 calibration points. Calibration is performed over RS485 by placing the electrode in a standard solution and writing the matching standard code to the calibration register (0x0043).
Standard Solution Codes
| Standard | Code |
|---|---|
| 0.1 ppm | 0x0002 |
| 1 ppm | 0x0004 |
| 10 ppm | 0x0008 |
| 100 ppm | 0x0010 |
| 1000 ppm | 0x0020 |
Calibration Procedure
- Place the electrode in the standard solution.
- Write the standard code to register 0x0043. Example (1 ppm point):
01 06 00 43 00 04 79 DD
- Poll register 0x0043 to read calibration status:
| Value | Meaning |
|---|---|
| 0 | Success (returned to measurement mode) |
| 1 | In progress (wait and re-read) |
| 2 | Incorrect standard solution value |
| 3 | Signal unstable or out of range after 180s |
| 4 | Sensor performance out of allowed range |
- Repeat for additional points.
Clear Calibration
Write 0x7FFF to register 0x0043 to clear all calibration data.
Factory Reset
Enter setup mode, then restore defaults:
01 06 00 40 00 50 88 22 (enter setup mode)
01 06 00 41 7F FF B9 AE (restore defaults and restart)
Revision History
| Rev | Date | Changes |
|---|---|---|
| 1.0 | June 2026 | Initial release |
| 1.1 | August 2026 | Layout revision |
| 1.2 | October 2026 | New layout; corrected Modbus example CRC values |