# AGRINOVO-SS-130: Self-cleaning Suspended Solids (SS) Sensor

Agrinovo technical datasheet, revision 1.2 (October 2026).
PDF version: https://agrinovo.io/datasheets/sensors/AGRINOVO-SS-130_Datasheet.pdf

## 1. Overview

The AGRINOVO-SS-130 is a self-cleaning digital suspended solids sensor based on infrared scattered light technology. The infrared light emitted by the source scatters when passing through the sample, and the scattered light intensity is proportional to the suspended solids concentration. A detector positioned at 135° receives the scattered light, and the sensor calculates the SS concentration by analyzing the signal strength.

### Key Features

- Measurement range: 0 to 50000 mg/L
- Resolution: 0.1 mg/L / 1 mg/L
- Accuracy: ±10% or 10 mg/L (whichever is greater)
- Repeatability: ±3%
- Self-cleaning wiper (optional)
- RS485 Modbus-RTU output
- IP68 / NEMA6P protection
- SUS316L stainless steel body

### Applications

- Wastewater treatment plant process monitoring
- Industrial process water monitoring
- Effluent discharge monitoring
- Environmental water quality monitoring

## 2. Specifications

| Parameter | Specification |
|:----------|:--------------|
| **Range** | 0 to 50000 mg/L (customizable) |
| **Accuracy** | ±10% of reading or 10 mg/L, whichever is greater |
| **Repeatability** | ±3% |
| **Resolution** | 0.1 mg/L or 1 mg/L (depending on range) |
| **Pressure Rating** | ≤0.2 MPa |
| **Calibration** | Standard solution / sample calibration |
| **Materials** | Body: SUS316L; End caps: POM; Cable: PUR |
| **Supply Voltage** | 12-24 VDC |
| **Protocol** | Modbus RS485 |
| **Storage Temp** | −15 to 50°C |
| **Operating Temp** | 0 to 45°C (non-freezing) |
| **Dimensions** | Without wiper: 50mm dia. × 205mm, NPT3/4 thread |
|  | With wiper: 50mm dia. × 215mm, NPT3/4 thread |
| **Weight** | 0.8 kg |
| **Protection** | IP68 / NEMA6P |
| **Cable Length** | 10 m standard, extendable to 100 m |

## 3. Wiring

| Wire Color | Function | Description |
|:-----------|:---------|:------------|
| **Brown** | V+ | Power Supply (12-24V DC) |
| **Black** | GND | Power Ground |
| **Blue** | RS485-A | Data+ |
| **White** | RS485-B | Data− |

## 4. Communication Settings

| Parameter | Value |
|:----------|:------|
| **Protocol** | Modbus-RTU |
| **Baud Rate** | 9600 bps (configurable: 4800/9600/19200/38400) |
| **Data Bits** | 8 |
| **Parity** | None |
| **Stop Bits** | 1 |
| **Default Address** | 0x01 (range 1 to 200) |

## 5. Register Map

### Measurement Registers (Function 0x03)

| Address | Description | Type | Length | R/W | Notes |
|:--------|:------------|:-----|:-------|:----|:------|
| 0x0000 | SS Value | Float | 2 | R | 0.0 to 50000 mg/L |
| 0x0002 | Internal Temperature | Float | 2 | R | |
| 0x0004 | SS Slope | Float | 2 | RW | 0.1 to 9.99, default 1.00 |
| 0x0006 | Sensor Signal | Float | 2 | R | |
| 0x0008 | SS Offset | Float | 2 | RW | ±50000, default 0.00 |

### Configuration Registers (Function 0x03 read / 0x06 write)

| Address | Description | Type | Length | R/W | Notes |
|:--------|:------------|:-----|:-------|:----|:------|
| 0x000B | Wiper Time | Int | 1 | R | Written via address 0x0015 |
| 0x000C | Response Time | Int | 1 | RW | 3 to 60 seconds |
| 0x000D | Electrode Type | Int | 1 | R | 1 = with wiper, 0 = without |
| 0x0010 | Baud Rate | Int | 1 | RW | 0=4800, 1=9600, 2=19200, 3=38400 |
| 0x0011 | Slave Address | Int | 1 | RW | 1 to 200 |
| 0x0014 | Manual Wipe | Int | 1 | W | Send 66 to trigger |
| 0x0015 | Auto Wipe Interval | Int | 1 | W | Interval in minutes |
| 0x00C8 | Factory Reset | Int | 1 | W | Send 60 to restore |

## 6. Reading Data

Read SS value and temperature (4 float registers from 0x0000):

```
Request:  01 03 00 00 00 04 44 09
Response: 01 03 08 XX XX XX XX XX XX XX XX [CRC]
```

**Decoding:**

| Register | Bytes | Format | Result |
|:---------|:------|:-------|:-------|
| 0x0000 to 0x0001 SS Value | 4 bytes | IEEE-754 Float32 | **mg/L** |
| 0x0002 to 0x0003 Temperature | 4 bytes | IEEE-754 Float32 | **°C** |

## 7. Address Configuration

### Change Address (0x01 → 0x02)

```
01 06 00 11 00 02 58 0E
```

### Change Baud Rate (9600 → 19200)

```
01 06 00 10 00 02 09 CE
```

### Broadcast Discovery

Use address 0xFE with only one device connected.

## 8. Calibration

The sensor is factory-calibrated. Two calibration methods are available: **slope correction** (single-point) and **multi-point correction**.

### Slope Correction (Single-Point)

1. Connect sensor to PC via RS485-to-USB adapter.
2. Clean the sensor lens thoroughly.
3. Immerse sensor slowly in SS standard solution.
4. Wait for reading to stabilize and record the value.
5. Calculate slope: **Slope = Standard Value / Measured Value**
6. Write the calculated slope to register 0x0004 (Float).

### Multi-Point Correction

| Step | Address | Type | Procedure |
|:-----|:--------|:-----|:----------|
| 1 | 0x0064 | Float | Place in pure water, write 0.01. Wait 20s. |
| 2 | 0x0080 | Float | Place in 100 mg/L standard, write 100. Wait 20s. |
| 3 | 0x0082 | Float | Place in 500 mg/L standard, write 500. Wait 20s. |
| 4 | 0x0084 | Float | Place in 1000 mg/L standard, write 1000. Wait 20s. |

### Calibration Notes

- Keep probe lens at least 5 cm from the bottom of the calibration cup
- Ensure no air bubbles on the lens surface
- Shield the calibration cup from ambient light during calibration

## 9. Installation Notes

**Placement**

- Use railing-mount bracket with DN32 pipe
- Submerge sensor fully in the measurement medium
- Avoid areas with excessive air bubbles or turbulence

**Maintenance**

- Clean the two optical lenses periodically
- Rinse with clean water, then wipe with detergent for stubborn deposits
- Inspect for damage; replace sealing rings annually
- Do NOT manually rotate the self-cleaning wiper  -  motor-driven only

## Ordering Information

| Part number | Description |
|:------------|:------------|
| AGRINOVO-SS-130 | Self-cleaning Suspended Solids (SS) Sensor |

## Revision History

| Rev | Date | Changes |
|:----|:-----|:--------|
| 1.0 | June 2026 | Initial release |
| 1.1 | August 2026 | Layout revision |
| 1.2 | October 2026 | New layout |

## Contact

Agrinovo, https://agrinovo.io/products, contact@agrinovo.io, +972 54 688 8148 (direct / WhatsApp).
Specifications subject to change without notice.