# AGRINOVO-PH-S100: Soil pH Sensor

Agrinovo technical datasheet, revision 1.0 (November 2026).
PDF version: https://agrinovo.io/datasheets/sensors/AGRINOVO-PH-S100_Datasheet.pdf

## 1. Overview

The AGRINOVO-PH-S100 is a spear-tip pH sensor that measures pH directly in soil, growing media and other semi-solid materials. A hardened piercing tip protects the glass electrode while the probe is pushed into the ground, and the transmitter is built into the probe, so it delivers a ready-to-use digital or 4 to 20 mA signal with temperature compensation and no separate pH meter.

### Key Features

- Direct in-soil measurement
- Piercing spear tip, 13 mm shaft
- pH range: 0 to 14.00
- Built-in temperature compensation
- RS485 Modbus-RTU or 4 to 20 mA
- Isolated power and output
- 3-point calibration
- Waterproof, PG13.5 mounting thread

### Applications

- Open-field and orchard soil
- Greenhouse beds and substrates
- Nurseries and potting media
- Compost and organic amendments
- Soil research and trial plots

## 2. Specifications

| Parameter | Specification |
|:----------|:--------------|
| **Measurement Range** | 0.00 to 14.00 pH |
| **Resolution** | 0.01 pH |
| **Accuracy** | ±0.1 pH |
| **Temperature Range** | 0 to 60°C |
| **Temperature Resolution** | 0.1°C |
| **Temperature Accuracy** | ±0.5°C |
| **Temperature Compensation** | Automatic or manual |
| **Calibration** | Up to 3 points (4.01 / 6.86 / 9.18 or 4.00 / 7.00 / 10.01) |
| **Output** | RS485 Modbus-RTU or 4 to 20 mA, selected when ordering |
| **Isolation** | 2500 Vrms, power and output |
| **Power Consumption** | About 0.6 W |

## 3. Physical Dimensions

| Parameter | Dimension |
|:----------|:----------|
| **Total Length** | 240 mm |
| **Insertion Shaft Diameter** | 13 mm |
| **Head Diameter** | 23 mm |
| **Mounting Thread** | PG13.5 |
| **Tip** | Piercing spear tip with protected glass electrode |

## 4. Electrical Characteristics

Each probe is built with one output type, selected when it is ordered.

| Output Type | Supply Voltage | Signal |
|:------------|:---------------|:-------|
| RS485 | 9 to 27 VDC | Modbus-RTU, pH, electrode mV and temperature |
| 4 to 20 mA | 24 VDC | 4.00 to 20.00 mA, resolution 0.01 mA, accuracy ±0.02 mA |

### Current to pH

| Output | Formula |
|:-------|:--------|
| 4 to 20 mA | pH = (mA - 4) × 0.875 |

## 5. Wiring (RS485)

| Wire Color | Function | Description |
|:-----------|:---------|:------------|
| **Red** | V+ | Power Supply (9 to 27 VDC) |
| **Black** | GND | Power Ground |
| **Green** | RS485-A | Data+ |
| **White** | RS485-B | Data- |

## 6. Communication Settings

| Parameter | Value |
|:----------|:------|
| **Protocol** | Modbus-RTU |
| **Baud Rate** | 9600 bps (1200 to 19200 selectable) |
| **Data Bits** | 8 |
| **Parity** | None |
| **Stop Bits** | 1 |
| **Default Address** | 0x01 |

## 7. Register Map

### Measurement Registers, Integer (Function 0x04)

| Address | Description | Scaling |
|:--------|:------------|:--------|
| 0x0000 | pH Value | Value ÷ 100 |
| 0x0001 | pH decimals and unit code | High byte decimals, low byte unit |
| 0x0002 | Electrode Signal (mV) | Direct, -1000 to 1000 |
| 0x0003 | Signal decimals and unit code | High byte decimals, low byte unit |
| 0x0008 | Temperature | Value ÷ 10 |
| 0x0009 | Temperature decimals and unit code | High byte decimals, low byte unit |

A value of 0x7FFF means over range and 0x8000 means under range.

### Measurement Registers, Float (Function 0x03)

| Address | Description | Format |
|:--------|:------------|:-------|
| 0x0000 to 0x0001 | pH Value | IEEE-754 float32, low word first |
| 0x0002 to 0x0003 | Electrode Signal (mV) | IEEE-754 float32, low word first |
| 0x0008 to 0x0009 | Temperature (°C) | IEEE-754 float32, low word first |

### Configuration Registers (Read 0x03, Write 0x06)

| Address | Description | Notes |
|:--------|:------------|:------|
| 0x0019 | Calibration Status | Read only. Bit 1 low point, bit 2 mid point, bit 3 high point |
| 0x001C | Acid Slope | Read only. Value ÷ 10, in % |
| 0x001D | Alkaline Slope | Read only. Value ÷ 10, in % |
| 0x001E | Slave Address | 1 to 247 |
| 0x001F | Baud Rate | 0 = 1200, 1 = 2400, 2 = 4800, 3 = 9600, 4 = 19200 |
| 0x0020 | Temperature Compensation | 0 = manual, 1 = automatic |
| 0x0021 | Temperature Setting | Value × 10. Offset in automatic mode, fixed temperature in manual mode |
| 0x0024 | Buffer Set | 0 = 4.01 / 6.86 / 9.18, 1 = 4.00 / 7.00 / 10.01 |
| 0x0043 | Calibration Command and Result | See Calibration |

## 8. Reading Data

Read 10 registers from 0x0000 with function 0x04:

```
Request:  01 04 00 00 00 0A 70 0D
Response: 01 04 14 02 BC 02 0A 00 00 00 00 00 00 00 00 00 00 00 00 00 FA 01 0B 7E 94
```

**Decoding:**

| Register | Hex | Decimal | Scaling | Result |
|:---------|:----|:--------|:--------|:-------|
| pH | 0x02BC | 700 | ÷100 | **7.00 pH** |
| Electrode Signal | 0x0000 | 0 | Direct | **0 mV** |
| Temperature | 0x00FA | 250 | ÷10 | **25.0°C** |

## 9. Calibration

Calibrate in buffer solutions before first use and at regular intervals. Always calibrate the mid point first; calibrating the mid point clears the earlier calibration.

| Step | Buffer | Command |
|:-----|:-------|:--------|
| 1. Mid point | 6.86 or 7.00 pH | `01 06 00 43 00 04 79 DD` |
| 2. Low point | 4.00 or 4.01 pH | `01 06 00 43 00 02 F9 DF` |
| 3. High point | 9.18 or 10.01 pH | `01 06 00 43 00 08 79 D8` |

### Calibration Procedure

1. Rinse the tip with distilled water and place it in the buffer
2. Send the command for that buffer
3. Read the result until it is no longer 1

```
Request:  01 03 00 43 00 01 75 DE
Response: 01 03 02 00 00 B8 44
```

| Result | Meaning |
|:-------|:--------|
| 0 | Calibration saved |
| 1 | Calibration in progress |
| 2 | Buffer not recognised within 180 s |
| 3 | Signal not stable within 180 s |
| 4 | Electrode slope or offset out of limits |

## 10. Address Configuration

### Change Address (0x01 → 0x02)

```
01 06 00 1E 00 02 68 0D
```

### Restore Default Settings

Send both commands in order. Calibration is cleared and the sensor restarts.

```
01 06 00 40 00 50 88 22
01 06 00 41 7F FF B9 AE
```

## 11. Installation Notes

**Measuring in Soil**

- Measure in moist soil; wet dry soil with distilled water first
- Push the probe in straight, without twisting
- Loosen hard or stony ground before inserting
- Press the soil around the shaft for full contact

**Sensor Care**

- Rinse the tip with clean water after use
- Keep the tip moist in storage, never dry
- Do not strike or lever on the tip
- Do not hot-plug

## Ordering Information

| Part number | Description |
|:------------|:------------|
| AGRINOVO-PH-S100 | Soil pH Sensor |

## Revision History

| Rev | Date | Changes |
|:----|:-----|:--------|
| 1.0 | November 2026 | Initial release |